In vivo volumetric and linear wear measurement of reverse shoulder arthroplasty at minimum 5-year follow-up.
Аннотация
BACKGROUND
Reverse shoulder arthroplasty is quickly becoming the most frequently performed glenohumeral joint replacement. The purpose of this study was to evaluate the volumetric and linear wear rates of ultrahigh-molecular-weight polyethylene humeral liners in vivo at a minimum 5-year follow-up.
METHODS
Radiostereometric analysis was used to image 15 patients at terminal range of motion in forward flexion, abduction, external rotation, and internal rotation and with the arm at the side. The relative position and orientation of the glenosphere and polyethylene were identified for each arm position. The apparent intersection of the glenosphere into the polyethylene was recorded as wear. Mean volumetric and linear wear rates were recorded, and Pearson correlation coefficients were applied to the 36-mm liners to assess the relationship between the wear rate and term of service.
RESULTS
The mean reverse shoulder arthroplasty term of service at the time of imaging was 8 ± 1 years (range, 6-11 years). The mean volumetric and linear wear rates for the 36-mm liners (n = 13) were 42 ± 22 mm3/yr (r = 0.688, P = .009) and 0.11 ± 0.03 mm/yr (r = 0.767, P = .002), respectively. The mean volumetric and linear wear rates for the 42-mm liners (n = 2) were 114 ± 44 mm3/yr and 0.17 ± 0.01 mm/yr, respectively. No single arm position was able to capture all recorded wear individually.
CONCLUSION
This study showed volumetric and linear wear rates of approximately 40 mm3/yr and 0.1 mm/yr, respectively, for the 36-mm polyethylene liners. The 42-mm liners showed higher wear rates, although a greater number of subjects is required for conclusive results. In vivo wear of reverse total shoulder arthroplasty is multidirectional and perceptible.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.