Disinfection of intraoral photostimulable phosphor plates with 70% ethyl alcohol and 70% isopropyl alcohol alters image contrast and produces image artifacts.
Аннотация
OBJECTIVE
This study evaluated changes in image quality and the presence and size of artifacts on radiographs obtained with photostimulable phosphor (PSP) plates after disinfection with 70% ethyl alcohol and 70% isopropyl alcohol.
STUDY
DESIGN
Eight PSP receptors (4 Express and 4 VistaScan) were used for the investigation. Two Express and 2 VistaScan receptors were disinfected with 70% ethyl alcohol, and 2 Express and 2 VistaScan receptors were disinfected with 70% isopropyl alcohol. Before disinfection and after every 5 disinfections, 2 radiographs of an aluminum stepwedge and 2 radiographs of an acrylic block were acquired using each receptor. The gray values of the radiographs were analyzed regarding contrast, density, uniformity, and noise. Four oral radiologists evaluated the images for the presence of artifacts, and 2 measured the artifacts.
RESULTS
Differences in image contrast were significantly larger with isopropyl alcohol (P < .0001). Artifact size was significantly greater with the VistaScan system when isopropyl alcohol was used as the disinfectant (P < .0001). The correlation between the number of disinfections and artifact size was moderately positive with the Express system (P < .05) and strongly positive with the VistaScan system (P < .05) with both disinfectants.
CONCLUSIONS
Small artifacts restricted to the edges of the image were observed with disinfection using both types of alcohol, and a positive correlation was found between artifact size and the number of disinfections. However, artifacts appeared earlier and of greater size with the VistaScan system when disinfected with 70% isopropyl alcohol. Thus, we recommend using 70% ethyl alcohol to disinfect PSP receptors, with attention placed on its action and evaporation times.