Micra transcatheter pacing system implantation via the right internal jugular vein: Preliminary results from a Polish multicenter registry.
Аннотация
BACKGROUND
Leadless pacemaker implantation via the right internal jugular vein (RIJV) is anemerging alternative to the traditional femoral approach, offering favorable anatomical access andpotential procedural advantages.
AIMS: Our aim was to assess the electrical performance, procedural efficiency and safety of Micraleadless pacemaker implantation via RIJV.
METHODS
Thirty-five consecutive patients who underwent Micra transcatheter pacing system (TPS) implantation via RIJV at 7 clinical centers in Poland between November 2024 and November2025 were included in this registry. The procedural safety and efficacy were assessed, and in-hospitalas well as short-term follow-up electrical parameters were analyzed.
RESULTS
Implantation success was achieved in 100% of cases. There were no significant changesin R-wave sensing (8.41 [2.45] vs. 9.20 [3.43] vs. 9.34 [3.60] mV; P = 0.45) or pacing threshold(0.5 [0.38-0.57] vs. 0.5 [0.38-0.63] vs. 0.5 [0.38-0.69] V; P = 0.25) between the procedure, predischarge,and 1-3-month follow-up. Impedance showed a significant decrease for a decrease across all timepoints (680 [620-745] vs. 630 [590-690] vs. 570 [535-607.5] Ω; P = 0.006). No mild or severe acuteor in-hospital complications related to the procedure were observed.
CONCLUSIONS
Micra TPS implantation via the RIJV is a highly effective and safe alternative to femoralaccess when performed in experienced cardiology centers.