Performance of Vitek MS v3.0 for Identification of Mycobacterium Species from Patient Samples by Use of Automated Liquid Medium Systems.

Miller Eric, Cantrell Christopher, Beard Melodie, Derylak Andrew, Babady N Esther, McMillen Tracy, Miranda Edwin, Body Barbara, Tang Yi-Wei, Vasireddy Ravikiran, Vasireddy Sruthi, Smith Terry, Iakhiaeva Elena, Wallace Richard J, Brown-Elliott Barbara A, Moreno Erik, Totty Heather, Deol Parampal

Сравнительное исследованиеОценочное исследованиеСтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

The accuracy and robustness of the Vitek MS v3.0 matrix-assisted laser desorption ionization-time of flight (MALDI-TOF) mass spectrometry (MS) system was evaluated by identifying mycobacteria from automated liquid-medium systems using patient samples. This is the first report to demonstrate that proteins within the liquid medium, its supplements, and decontamination reagents for nonsterile patient samples do not generate misidentification or false-positive results by use of the Vitek MS v3.0 system. Prior to testing with patient samples, a seeded-culture study was conducted to challenge the accuracy of the Vitek MS system at identifying mycobacteria from liquid medium by mimicking a clinical workflow. Seventy-seven Mycobacterium strains representing 21 species, seeded in simulated sputum, were decontaminated, inoculated into BacT/Alert MP liquid culture medium, incubated until positivity, and identified using the Vitek MS system. A total of 383 liquid cultures were tested, of which 379 (99%) were identified correctly to the species/complex/group level, 4 (1%) gave a "no-identification" result, and no misidentifications were observed. Following the simulated-sputum study, a total of 73 smear-positive liquid-medium cultures detected using BD BBL MGIT and VersaTREK Myco liquid media were identified by the Vitek MS system. Sixty-four cultures (87.7%) were correctly identified to the species/complex/group level; 7 (9.6%) resulted in no identification; and 2 (2.7%) were misidentified at the species level. These results indicate that the Vitek MS v3.0 system is an accurate tool for routine diagnostics of Mycobacterium species isolated from liquid cultures.

Для цитирования: Miller Eric, Cantrell Christopher, Beard Melodie, Derylak Andrew, Babady N Esther, McMillen Tracy, Miranda Edwin, Body Barbara, Tang Yi-Wei, Vasireddy Ravikiran, Vasireddy Sruthi, Smith Terry, Iakhiaeva Elena, Wallace Richard J, Brown-Elliott Barbara A, Moreno Erik, Totty Heather, Deol Parampal. Performance of Vitek MS v3.0 for Identification of Mycobacterium Species from Patient Samples by Use of Automated Liquid Medium Systems.. Journal of clinical microbiology. 2018. DOI: 10.1128/jcm.00219-18.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.