Displacement of intraorbital ferromagnetic foreign bodies induced by magnetic resonance imaging: Quantification using an animal model.

Cathelineau Camille, Ghemame Marwane, Le Boëdec Antoine, Carsin-Nicol Béatrice, Saint-Jalmes Hervé, Éliat Pierre-Antoine, Mouriaux Frédéric, Ferré Jean-Christophe

СтатьяПеревод в очереди

Аннотация

RATIONALE AND

OBJECTIVES

The presence of ferromagnetic foreign bodies in the intraorbital area, eyelids, or intraorbital fat of an examined subject represents a contraindication to magnetic resonance imaging (MRI). We sought to measure the displacement of one or two intraocular and intraorbital ferromagnetic foreign bodies in ex vivo porcine heads following 1.5 T MRI scan compared to 5-min walk test.

MATERIALS AND

METHODS

In this ex-vivo controlled laboratory study, a total of 48 1-, 2-, and 4-mm steel balls were surgically implanted into suprachoroidal, intraorbital fat, intrapalpebral, and intravitreal locations of 36 fresh porcine heads. Ferromagnetic foreign bodies displacement was measured by comparing computed tomography scan before and after both a 5-min walk test to simulate handling-related movement and 1.5 T MRI scan. Comparison of groups was performed using the non-parametric Mann-Whitney-Wilcoxon test.

RESULTS

Global median displacement after 5-min walk test was 0.8 mm. Compared to the control group, the ferromagnetic foreign bodies displacement after MRI was significant for the suprachoroidal location with single ferromagnetic foreign body (2.3 mm, p = 0.0282), and for the intravitreal location with single ferromagnetic foreign body (4.5 mm, p = 0.0282) and double ferromagnetic foreign bodies (5.0 mm, p = 0.0282). Compared to the control group, ferromagnetic foreign body displacement after MRI was significant for the double 2-mm ferromagnetic foreign bodies (2.2 mm, p = 0.0130), single 4-mm ferromagnetic foreign body (4.9 mm, p = 0.0107), and double 4-mm ferromagnetic foreign bodies (9.0 mm, p = 0.0084).

CONCLUSION

Ferromagnetic foreign body displacements after 1.5 T magnetic field exposition were not significant for extraocular intrapalpebral and intraorbital fat locations. Neither were there significant displacements for 1-mm ferromagnetic foreign body, irrespective of the location. On the other hand, the presence of two objects increased the displacement of 2- and 4-mm ferromagnetic foreign bodies.

Для цитирования: Cathelineau Camille, Ghemame Marwane, Le Boëdec Antoine, Carsin-Nicol Béatrice, Saint-Jalmes Hervé, Éliat Pierre-Antoine, Mouriaux Frédéric, Ferré Jean-Christophe. Displacement of intraorbital ferromagnetic foreign bodies induced by magnetic resonance imaging: Quantification using an animal model.. Research in diagnostic and interventional imaging. 2026. DOI: 10.1016/j.redii.2025.100064.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.