MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jse.2014.04.005

Медицинская библиотека · Journal of shoulder and elbow surgery · 2014

The TESS reverse shoulder arthroplasty without a stem in the treatment of cuff-deficient shoulder conditions: clinical and radiographic results.

Teissier Philippe, Teissier Jacques, Kouyoumdjian Pascal, Asencio Gérard

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Reverse total shoulder arthroplasty (RSA) is a recent concept that enables good functional outcomes in cases of massive rotator cuff tear and cuff tear arthropathy. Design parameters influence the functional results and complications. The purpose of this study is to present the results of a novel RSA, the Total Evolutive Shoulder System (TESS; Biomet, Warsaw, IN, USA), based on a reverse corolla without a stem.

METHODS

We enrolled 101 patients with 105 RSAs in a prospective study, with a minimum follow-up period of 24 months. The analysis concerned 91 RSAs in 87 patients (61 men and 26 women), with a mean age of 73 years, at a mean follow-up of 41 months (range, 24-69 months).

RESULTS

Ninety-six percent of patients rated their satisfaction as good or excellent. Mean flexion was 143° (range, 90°-170°), and mean external rotation was 39° (range, 20°-70°). The Constant score improved from 40 points preoperatively to 68 points at last follow-up (P < .001). The mean American Shoulder and Elbow Surgeons score was 24 points. The mean neck-shaft angle was 154° (range, 142°-165°). Inferior scapular notching occurred in 17 cases (19%). The notching rate was higher when the glenometaphyseal angle increased (P < .001), when the inferior tilt decreased (P = .003), and when the neck-shaft angle increased. There was no evidence of component loosening.

CONCLUSION

TESS RSA provided encouraging midterm results with favorable outcomes and a low rate of complications. The stemless TESS with a reverse corolla is a reliable, less invasive system.

Для цитирования: Teissier Philippe, Teissier Jacques, Kouyoumdjian Pascal, Asencio Gérard. The TESS reverse shoulder arthroplasty without a stem in the treatment of cuff-deficient shoulder conditions: clinical and radiographic results.. Journal of shoulder and elbow surgery. 2014. DOI: 10.1016/j.jse.2014.04.005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.