MedElevator Войти в исследования
← Поиск публикаций

DOI 10.14444/8652

Медицинская библиотека · Il Giornale di chirurgia · 2018

Concomitant small cell carcinoma and adenocarcinoma of the gallbladder: a case report.

Kazarian Gregory S, Jordan Yusef J, Johnson Mitchell, Bhargava Satyaj, Cecere Robert, Hirase Takashi, Qureshi Sheeraz, Dowdell James, Sheha Evan, Lovecchio Francis, Iyer Sravisht

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Interspinous process devices (IPDs) introduce a new class of complications to surgical decompression without fusion: hardware-related complications. The purpose of this study was to describe the adverse events associated with IPDs.

STUDY

DESIGN

This was a retrospective review of the Food and Drug Administration Manufacturer and User Facility Device Experience database.

METHODS

The database was queried from its inception to November 2022 for reports associated with "Prosthesis, Spinous Process Spacer/Plate." Entries were categorized by event type, patient impact, and interventions.

RESULTS

A total of 943 surgery-related adverse events were identified. The most common intraoperative events were implant malfunctions (39.7%, n = 374) and fractures (2.2%, n = 21). The most common postoperative events were persistent pain (26.6%, n = 251), implant migration (19.1%, n = 180), and fracture (6.8%, n = 64). The most common resultant outcome of an adverse event was the need for revision surgery (48.8%, n = 460). The need for revision surgery was common in patients who experienced fracture (47.1%), implant migration (84.5%), infection (76.7%), and neurological complications (76.9%). Implant migration, fracture, and implant malfunction, 3 complications that are unique to decompression with an IPD as compared with traditional laminectomy, accounted for 45.9% of revisions (211/460), and revision was required in 33.0% of cases where 1 of these complications was reported (211/640). Implant malfunction made up 21.2% of Coflex complications, 47.3% of Superion complications, and 5.2% of X-Stop complications.

CONCLUSIONS

The most common adverse events were implant malfunction, inadequate efficacy, implant migration, and fracture. Concerningly, these complications require revision surgery in one-third of cases when they occur. Implant-specific assessments demonstrate a high prevalence of implant malfunctions for the Coflex and Superion implants.

CLINICAL RELEVANCE: Interspinous process devices introduce a new class of complications to isolated spinal decompression surgery: implant-related complications. These complications occur both intraoperatively and postoperatively, and they frequently necessitate revision surgery.

Для цитирования: Kazarian Gregory S, Jordan Yusef J, Johnson Mitchell, Bhargava Satyaj, Cecere Robert, Hirase Takashi, Qureshi Sheeraz, Dowdell James, Sheha Evan, Lovecchio Francis, Iyer Sravisht. Concomitant small cell carcinoma and adenocarcinoma of the gallbladder: a case report.. Il Giornale di chirurgia. 2018. DOI: 10.14444/8652.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.