MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s10396-024-01485-1

Медицинская библиотека · Journal of medical ultrasonics (2001) · 2024

Usefulness of color Doppler and strain elastography adjunctive to B-mode ultrasonography in the diagnosis of non-mass abnormalities of the breast: results of the BC-07 multicenter study of 385 cases.

Okuno Toshitaka, Watanabe Takanori, Yamaguchi Takuhiro, Konno Sachiyo, Takaki Rie, Watanabe Ryoji, Ban Kanako, Hirokaga Koichi, Tsuruoka Masahiko, Morita Takako

СтатьяМногоцентровое исследованиеПеревод в очереди

Аннотация

PURPOSE: The concept of non-mass abnormalities of the breast has been employed in Japan for approximately 20 years. Although B-mode findings are classified as non-mass abnormalities, the usefulness of adding color Doppler ultrasonography (US) and strain elastography to B-mode US is unclear. Therefore, we conducted a multicenter study (JABTS BC-07) to establish the diagnostic criteria for breast US, including color Doppler and elastography, for non-mass abnormalities of the breast and verify their diagnostic usefulness.

METHODS

We registered US images of non-mass abnormalities of the breast and their clinical and histopathological data from 13 institutions (202 malignant and 183 benign non-mass lesions). Furthermore, we evaluated the centralized image interpretation usefulness of the diagnostic criteria for B-mode and color Doppler US, as well as the sensitivity and specificity when color Doppler US and elastography were added to B-mode US.

RESULTS

Echogenic foci in the mammary gland (odds ratio 3.45, 95% confidence interval [CI] 1.92-6.19, p < 0.0001) and the configuration of internal solid components of the ducts (odds ratio 0.056, 95% CI 0.005-0.591, p < 0.0165) significantly differentiated benign and malignant non-mass abnormalities. The sensitivity of B-mode alone (83.7%) was significantly improved by adding color Doppler US (93.1%) (p = 0.0004); however, adding color Doppler US and elastography to B-mode US made no significant difference in either sensitivity or specificity.

CONCLUSION

Although adding color Doppler US and elastography to B-mode US improved sensitivity, the diagnostic significance was limited. Therefore, a comprehensive diagnostic method comprising mammography and magnetic resonance imaging is warranted.

Для цитирования: Okuno Toshitaka, Watanabe Takanori, Yamaguchi Takuhiro, Konno Sachiyo, Takaki Rie, Watanabe Ryoji, Ban Kanako, Hirokaga Koichi, Tsuruoka Masahiko, Morita Takako. Usefulness of color Doppler and strain elastography adjunctive to B-mode ultrasonography in the diagnosis of non-mass abnormalities of the breast: results of the BC-07 multicenter study of 385 cases.. Journal of medical ultrasonics (2001). 2024. DOI: 10.1007/s10396-024-01485-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.