MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/1120672120977820

Медицинская библиотека · European journal of ophthalmology · 2020

The influence of dead spaces and the designs of injectors on the amount of drug dose in intra-vitreal injection.

Havuz Erol, Güdül Havuz Seda, Gokmen Onur

СтатьяПеревод в очереди

Аннотация

AIM: To evaluate the dead spaces resulting from different designs of the insulin injectors used for intravitreal injections and the amounts of drug doses.

METHODS

In the study, five different brands of sterile insulin injectors of 1 mL were used for the test. The weight of the injectors was determined before and after filling the injectors with 0.05 and 0.1 mL distilled water. The weight of the injectors was measured with and without the needle after the water within the injectors had been taken out and weight differences were measured. The difference between the intended amount of fluid to be thrown out and the weight of the fluid remaining in the injector was calculated as percent error.

RESULTS

After throwing out 0.05 mL distilled water from the injector, weights of the Beybi®, Traf®, Becton Dickinson®, Ayset®, and Setojet® brands of injectors with 30 G needle were detected to increase the mean 0.0220 ± 0.006 g, 0.0208 ± 0.008 g, 0.0355 ± 0.016 g, 0.0219 ± 0.017 g, and 0.0150 ± 0.007 g, respectively compared to the weights of the dry injectors. The 0.1 mL injector group was found to be mean 0.0350 ± 0.014 g, 0.0264 ± 0.008 g, 0.0405 ± 0.015 g, 0.0272 ± 0.013 g, and 0.0245 ± 0.014 g, respectively. The maximum increase due to the dead spaces in the injectors was found in the Becton Dickinson® injector, both in the 0.05 mL and the 0.1 mL groups (p < 0.000).

CONCLUSION

The injector designs may affect the dose of IVI required to be given. More correct amounts of drugs may be administered via the intra-vitreal route through designs that reduce the dead spaces at the end part of the injector and between the inner wall and the plunger.

Для цитирования: Havuz Erol, Güdül Havuz Seda, Gokmen Onur. The influence of dead spaces and the designs of injectors on the amount of drug dose in intra-vitreal injection.. European journal of ophthalmology. 2020. DOI: 10.1177/1120672120977820.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.