Clinical Accuracy of the G.LAB MD41A0 Upper Arm Blood Pressure Monitor in Pregnancy and Pre-Eclampsia Using the AAMI/ESH/ISO as Reference Standard.

СтатьяВалидационное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Zhang Ying, Shi Ziyun, Liang Haixia, Chen Meixia, Sun Yan, Shang Fujun, Wan Yi

Аннотация

OBJECTIVE: Validation of blood pressure (BP) monitors in pregnant women are needed to ensure the accurate measurement of BP in pregnancy. Therefore the study aimed to evaluate the measurement accuracy of the G.LAB MD41A0 oscillometric automatic upper-arm blood pressure monitor in pregnancy and pre-eclampsia women according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018).

METHODS: A total of 45 pregnant women were included in the cross-sectional validation study. The same left-arm sequential method was used for blood pressure measurement, and the blood pressure differences between the test device and the mercury standard reference were assessed according to the AAMI/ESH/ISO Universal Standard.

RESULTS: The participants included 15 normotensives, 15 gestational hypertension and 15 pre-eclampsia. The average (mean ± SD) differences between the device and mercury standard was -0.84±3.88 mmHg and -0.58±3.35 mmHg for systolic and diastolic blood pressure, respectively, for the validation Criterion 1. The SD of the averaged differences between the device and reference readings per participant was 2.92 mmHg and 2.28 mmHg for systolic and diastolic blood pressure, respectively, for the Criterion 2 of the universal standard.

CONCLUSION: The G.LAB MD41A0 automatic upper-arm blood pressure monitor fulfills the criteria of the AAMI/ESH/ISO Universal Standard and can be recommended for clinical use and self-measurement in pregnancy and pre-eclampsia.