MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s11548-021-02365-6

Медицинская библиотека · International journal of computer assisted radiology and surgery · 2021

Learning effectiveness of using augmented reality technology in central venous access procedure: an experiment using phantom and head-mounted display.

Suzuki Kazufumi, Morita Satoru, Endo Kenji, Yamamoto Takahiro, Fujii Shuhei, Ohya Jun, Masamune Ken, Sakai Shuji

СтатьяПеревод в очереди

Аннотация

PURPOSE: Augmented reality (AR) technology improves the learning process in interventional radiology. This study hypothesized that using AR to train for central venous access is superior to using ultrasound alone.

METHODS

This study used an AR central venous catheterization phantom with an internal jugular vein (IJV) and subclavian vein (SCV) made of resin body and soft tubing. Ten radiologists attempted to punctuate, using needle placement simulation, under three conditions (ultrasound-, augmented reality-, and ultrasound and AR-guided methods; US-only, AR-only, and US+AR, respectively) using a smart-glass device (HoloLens, Microsoft, Redmond, WA, USA). Subjective (anatomical understanding and self-confidence for procedure) and objective evaluations (optimized needle position and time) were recorded for each condition.

RESULTS

The subjective IJV evaluation showed no difference among the guiding methods (p = 0.26 and p = 0.07 for anatomical understanding and self-confidence for procedure, respectively). Conversely, there were significant improvements in subjective and objective evaluations for SCV using the AR-only and US+AR methods (p < 0.05) and US+AR method (p < 0.05), respectively. The AR-only method reduced the time required to fix the needle position to puncture the SCV (p < 0.05), but its objective evaluation did not improve compared with the US-only method (p = 0.20).

CONCLUSION

Adding the AR-guided method to the US-guided method improved subjective and objective evaluations in the SVC procedure. The AR technology-assisted training may be more beneficial for use in difficult procedures. Though the AR-only method saved time, no time saving is expected with AR+US method.

Для цитирования: Suzuki Kazufumi, Morita Satoru, Endo Kenji, Yamamoto Takahiro, Fujii Shuhei, Ohya Jun, Masamune Ken, Sakai Shuji. Learning effectiveness of using augmented reality technology in central venous access procedure: an experiment using phantom and head-mounted display.. International journal of computer assisted radiology and surgery. 2021. DOI: 10.1007/s11548-021-02365-6.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.