DOI 10.1016/j.jmr.2026.108081

Медицинская библиотека · Journal of magnetic resonance (San Diego, Calif. : 1997) · 2026

Low-power optimal control pulse sequences enhance the sensitivity of 5 mm triple resonance cryogenic probe optimized for proton-detection at 1.2 GHz NMR instruments.

Joseph David, Kümmerle Rainer, Freytag Nicolas, Griesinger Christian

СтатьяПеревод в очереди

Аннотация

Improving the sensitivity of nuclear magnetic resonance (NMR) spectroscopy requires advancements in both instrument technology and experimental methodology. In this study, we introduce a 5 mm triple resonance cryogenic probe optimized for proton-detection designed for 1.2 GHz instruments, leveraging optimal control pulse sequences to enhance performance. Our results show a 56% increase in average sensitivity for a benzene sample and up to a 50% improvement for a lossy concentration limited biological sample, reducing experimental time by more than a factor of two compared with the 3 mm cryogenic probes currently in use. Additionally, we systematically optimized the experimental conditions to fully exploit the capabilities of GHz-class magnets. The gain with the 5 mm cryogenic probe at 1.2 GHz was possible using low power optimal control pulses. Hence, we developed a library of optimal control triple resonance experiments, enabling boosted sensitivity for advanced NMR applications at 1.2 GHz instruments.

Для цитирования: Joseph David, Kümmerle Rainer, Freytag Nicolas, Griesinger Christian. Low-power optimal control pulse sequences enhance the sensitivity of 5 mm triple resonance cryogenic probe optimized for proton-detection at 1.2 GHz NMR instruments.. Journal of magnetic resonance (San Diego, Calif. : 1997). 2026. DOI: 10.1016/j.jmr.2026.108081.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.