MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/jcu.70154

Медицинская библиотека · Journal of clinical ultrasound : JCU · 2025

Assessing Acoustic Safety in Prenatal Ultrasound: A Comparative Study of Doppler and B-Mode Examinations.

Farzanegan Zahra, Fatahiasl Jafar, Arjmandi Maryam, Alboshokeh Seyed Ali, Mousavi Seyede Yasamin, Jafari Mostafa, Tahmasbi Marziyeh, Masihi Sara

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

OBJECTIVES

The widespread use of ultrasound in prenatal care raises concerns about fetal safety due to acoustic exposure. This study compared the thermal index (TI), mechanical index (MI), and scan duration between Doppler and B-mode examinations, while identifying the clinical indications for the use of Doppler.

METHODS

A cross-sectional study was conducted on 264 pregnancy ultrasound examinations performed at one educational hospital and two private clinics. Certified radiologists and gynecologists performed all scans. Data were collected using a structured checklist. Scan modality, gestational age, TI, MI, scan duration, and clinical indications were recorded. Convenience sampling was used. Statistical analysis was performed using SPSS version 26, with significance set at p < 0.05.

RESULTS

Doppler scans comprised 23.5% of all examinations and were primarily indicated for intrauterine growth restriction (IUGR) and maternal hypertension. TI values were significantly higher in Doppler scans, while MI values exceeded the recommended threshold (0.4) across both modalities, especially in B-mode scans. Scan durations were longer for B-mode during later gestational weeks. No significant differences in safety indices were observed across imaging centers or ultrasound systems.

CONCLUSION

Although TI values remained within recommended safety ranges, persistently elevated MI values emphasize the need for vigilance. Doppler should not be routinely restricted but reserved for clinically justified cases. Sonographers should actively monitor TI and MI values during scans and minimize exposure time, adjusting scan duration according to gestational age and modality to ensure fetal safety.

Для цитирования: Farzanegan Zahra, Fatahiasl Jafar, Arjmandi Maryam, Alboshokeh Seyed Ali, Mousavi Seyede Yasamin, Jafari Mostafa, Tahmasbi Marziyeh, Masihi Sara. Assessing Acoustic Safety in Prenatal Ultrasound: A Comparative Study of Doppler and B-Mode Examinations.. Journal of clinical ultrasound : JCU. 2025. DOI: 10.1002/jcu.70154.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.