MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00402-023-05142-z

Медицинская библиотека · Archives of orthopaedic and trauma surgery · 2024

Undersizing of the tibial component in Oxford unicompartmental knee arthroplasty (UKA) increases the risk of periprosthetic fractures.

Watrinet Julius, Blum Philipp, Maier Michael, Klingbeil Steffen, Regenbogen Stephan, Augat Peter, Schipp Rolf, Reng Wolfgang

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

Tibial periprosthetic fractures (TPF) after unicompartmental knee arthroplasty (UKA) are a rare condition that affects about 1% of cases. Known risk factors include age, sex, body mass index (BMI), and bone density, as well as surgical technique and prosthesis design. The purpose of the study was to determine if undersizing of the tibial component in relation to the femoral component increases the risk of tibial periprosthetic fractures.

MATERIAL AND

METHODS

Over a 6-year-period 1542 patients with cemented (n = 363) and uncemented (n = 1179) medial UKA were retrospectively evaluated. Tibial periprosthetic fractures were identified and classified, and epidemiologic data were documented at follow-up. Undersizing was defined as a smaller tibial component compared to the femoral implant. The association of potential risk factors for TPF with the incidence of TPF was investigated with binominal logistic regression.

RESULTS

Fourteen patients (0.9%) suffered from TPF at a median of 1 month after surgery. The mean follow-up period was 5.9 ± 1.7 years. Fractures were more common in cases with undersized tibial components [odds ratio (OR) 3.2, p < 0.05]. Furthermore, older age (OR 1.1, p < 0.05) and female sex (OR 6.5, p < 0.05) were identified as significant risk factors, while BMI (p = 0.8) and cemented implantation (p = 0.2) had no effect on fracture rate. Revision surgery included open reduction and internal fixation or conversion to total knee arthroplasty.

CONCLUSIONS

Undersizing of implant sizes in UKA increases the risk for TPF especially in patients with small tibial implants. Therefore, mismatched implants should be avoided for UKA particularly when risk factors like obesity, older age, or female gender are present. Tibial periprosthetic fractures were successfully treated by open reduction and internal fixation or conversion to total knee arthroplasty.

Для цитирования: Watrinet Julius, Blum Philipp, Maier Michael, Klingbeil Steffen, Regenbogen Stephan, Augat Peter, Schipp Rolf, Reng Wolfgang. Undersizing of the tibial component in Oxford unicompartmental knee arthroplasty (UKA) increases the risk of periprosthetic fractures.. Archives of orthopaedic and trauma surgery. 2024. DOI: 10.1007/s00402-023-05142-z.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.