Trileaflet Semilunar Valve Reconstruction: Acute Porcine in Vivo Evaluation.
Аннотация
Objective
The surgical treatment of malformed semilunar valves in congenital heart defects is challenging in terms of providing both longevity and the potential to grow with the recipient. We investigated a new surgical technique "Trileaflet Semilunar Valve Reconstruction" in an acute porcine model, a technique with geometrical properties that could remain sufficient and allow for some growth with the child.
Methods
An acute 60-kg porcine model was used. With echocardiography, baseline pulmonary valvular geometry and hemodynamics were investigated. On cardiopulmonary bypass, the pulmonary leaflets were explanted, and the Trileaflet Semilunar Valve Reconstruction was performed with customized homograft-treated pericardial neo-leaflets. Off bypass, hemodynamics was reassessed.
Results
Twelve animals were investigated. The neo-valves were found sufficient in ten animals and with minimal regurgitation in two animals. The neo-valve had a peak gradient of 3 ± 2 mm Hg with a peak velocity of 0.8 ± 0.2 m/s. The coaptation in the neo-valve had a mean increase of 4 ± 3 mm, P < .001. The neo-valve had a windmill shape in the echocardiographic short-axis view, and the neo-leaflets billowed at the annular plane in the long-axis view.
Conclusions
In this acute porcine model, the neo-valve had no clinically significant regurgitation or stenosis. The neo-valve had an increased coaptation, a windmill shape, and leaflets that billowed at the annular plane. These geometric findings may allow for sustained sufficiency as the annular and pulmonary artery dimension increase with the child's growth. Further long-term studies should be performed to evaluate the efficacy and the growth potential.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.