DOI 10.1111/apa.70622

Медицинская библиотека · Acta paediatrica (Oslo, Norway : 1992) · 2026

In Vitro Performance of Valved Holding Chambers in Paediatric Asthma Controller Therapy.

Ojanperä Laura, Lehtimäki Lauri, Csonka Péter

СтатьяПеревод в очереди

Аннотация

AIM: Research on the drug delivery of inhaled corticosteroids and combination medications via valved holding chambers (VHCs) with children's breathing patterns is limited. We aimed to evaluate the in vitro performance of commercially available VHCs without facemasks under preschooler breathing conditions, focusing on total delivered dose and fine particle dose (FPD).

METHODS

We evaluated the FPD in 1-5 μm range of fluticasone/salmeterol combination inhaler, fluticasone propionate mono-inhaler, and ciclesonide delivered through three VHCs: AeroChamber Plus Flow-Vu, EasyChamber, and OptiChamber Diamond. Drug delivery was measured using a Next Generation Impactor with simulated breathing representative of a 4-year-old child.

RESULTS

The total delivered dose compared to the label claim was highest for ciclesonide with all the VHCs. The median FPD in 1-5 μm range was significantly higher with EasyChamber compared to AeroChamber and OptiChamber for salmeterol and fluticasone from combination-inhaler and ciclesonide. Overall, relative throat deposition was lowest for ciclesonide and highest for fluticasone mono-inhaler. OptiChamber yielded lowest relative throat depositions for ciclesonide and fluticasone mono-inhaler, while EasyChamber gave lowest relative throat depositions for salmeterol and fluticasone from combination inhaler.

CONCLUSION

There were notable differences between the tested VHC's drug delivery. Ciclesonide exhibited favourable aerodynamic properties than fluticasone.

Для цитирования: Ojanperä Laura, Lehtimäki Lauri, Csonka Péter. In Vitro Performance of Valved Holding Chambers in Paediatric Asthma Controller Therapy.. Acta paediatrica (Oslo, Norway : 1992). 2026. DOI: 10.1111/apa.70622.