Heart rhythm O2 · 2021
Performance and outcomes of transvenous rotational lead extraction: Results from a prospective, monitored, international clinical study.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Sharma Saumya, Lee Byron K, Garg Anuj, Peyton Robert, Schuler Brian T, Mason Pamela, Delnoy Peter Paul, Gallagher Mark M, Hariharan Ramesh, Schaerf Raymond, Du Ruirui, Serratore Nina D, Starck Christoph T
Аннотация
BACKGROUND: Transvenous lead extraction (TLE) plays a critical role in managing patients with cardiovascular implantable electronic devices. Mechanical TLE tools, including rotational sheaths, are used to overcome fibrosis and calcification surrounding leads. Prospective clinical data are limited regarding the safety and effectiveness of use of mechanical TLE devices, especially rotational tools.
OBJECTIVE: To prospectively investigate the safety and effectiveness of mechanical TLE in real-world usage.
METHODS: Patients were enrolled at 10 sites in the United States and Europe to evaluate the use of mechanical TLE devices. Clinical success, complete procedural success, and complications were evaluated through follow-up (median, 29 days). Patient data were source verified and complications were adjudicated by an independent clinical events committee (CEC).
RESULTS: Between October 2018 and January 2020, mechanical TLE tools, including rotational sheaths, were used to extract 460 leads with a median indwell time of 7.4 years from 230 patients (mean age 64.3 ± 14.4 years). Noninfectious indications for TLE were more common than infectious indications (61.5% vs 38.5%, respectively). The extracted leads included 305 pacemaker leads (66.3%) and 155 implantable cardioverter-defibrillator leads (33.7%), including 85 leads with passive fixation (18.5%). A bidirectional rotational sheath was needed for 368 leads (88.0%). Clinical success was obtained in 98.7% of procedures; complete procedural success was achieved for 96.3% of leads. CEC-adjudicated device-related major complications occurred in 6 of 230 (2.6%) procedures. No isolated superior vena cava injury or procedural death occurred.
CONCLUSION: This prospective clinical study demonstrates that use of mechanical TLE tools, especially bidirectional rotational sheaths, are effective and safe.
OBJECTIVE: To prospectively investigate the safety and effectiveness of mechanical TLE in real-world usage.
METHODS: Patients were enrolled at 10 sites in the United States and Europe to evaluate the use of mechanical TLE devices. Clinical success, complete procedural success, and complications were evaluated through follow-up (median, 29 days). Patient data were source verified and complications were adjudicated by an independent clinical events committee (CEC).
RESULTS: Between October 2018 and January 2020, mechanical TLE tools, including rotational sheaths, were used to extract 460 leads with a median indwell time of 7.4 years from 230 patients (mean age 64.3 ± 14.4 years). Noninfectious indications for TLE were more common than infectious indications (61.5% vs 38.5%, respectively). The extracted leads included 305 pacemaker leads (66.3%) and 155 implantable cardioverter-defibrillator leads (33.7%), including 85 leads with passive fixation (18.5%). A bidirectional rotational sheath was needed for 368 leads (88.0%). Clinical success was obtained in 98.7% of procedures; complete procedural success was achieved for 96.3% of leads. CEC-adjudicated device-related major complications occurred in 6 of 230 (2.6%) procedures. No isolated superior vena cava injury or procedural death occurred.