Cephalic vein access by modified Seldinger technique for lead implantations.

Rademakers Leonard M, Bracke Frank A

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Venous access for cardiac implantable electronic devices (CIED) is commonly performed by cephalic venous cut down, or axillary or subclavian vein puncture. The latter technique carries a risk of complications such as pneumothorax or lead crush. Cephalic venous cut down is free of these complications but often less successful due to technical difficulties. We report a highly successful, simplified cephalic venous access with a modified Seldinger technique.

METHODS

We prospectively studied 221 patients undergoing de novo implantation of a one, two, or three lead device system performed over a 1-year period at our center, and assessed the efficacy of the cephalic vein access including the number of leads successfully placed for each procedure.

RESULTS

In 83% of patients undergoing CIED implantation, a suitable cephalic vein was present. In respectively 98% and 99% of patients undergoing single- or dual-chamber CIED implantation, lead placement could be performed exclusively via the cephalic vein and in 72% of cardiac resynchronization therapy implants, all three leads were placed via cephalic access.

CONCLUSION

A novel, technically simple cephalic venous catheterization technique provides high success rates for any CIED lead implantation.

Для цитирования: Rademakers Leonard M, Bracke Frank A. Cephalic vein access by modified Seldinger technique for lead implantations.. Pacing and clinical electrophysiology : PACE. 2021. DOI: 10.1111/pace.14200.