Adverse device-device interaction between pacemaker and subcutaneous implantable cardiac defibrillator.

Abbott Nicholas, Bender Aron, Henrikson Charles, Miller Jared, Nazer Babak, Balaji Seshadri

Клинический случайСтатьяПеревод в очереди

Аннотация

A 26-year-old patient with prior surgery for Ebstein's anomaly and a pacemaker (placed for post-surgical heart block and poor underlying rhythm) underwent SICD was placement. During defibrillation testing, device-device interaction led to undersensing of ventricular fibrillation with failure to shock. Increasing the pacemaker sensitivity resolved the problem but post shock pacing was unable to capture the heart after both shocks. The patient underwent removal of both the pacemaker and the SICD and placement of a transvenous ICD. Complex device-device interactions can occur in patients who are pacemaker dependent and undergo placement of a SICD.

Для цитирования: Abbott Nicholas, Bender Aron, Henrikson Charles, Miller Jared, Nazer Babak, Balaji Seshadri. Adverse device-device interaction between pacemaker and subcutaneous implantable cardiac defibrillator.. Pacing and clinical electrophysiology : PACE. 2021. DOI: 10.1111/pace.14325.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.