DOI 10.1097/mao.0000000000004747

Медицинская библиотека · Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2025

Spatiotemporal Dynamics of Cochlear Implant Electrode Impedance in Standard and Dexamethasone-Eluting Electrode Arrays.

Khan Uzair A, Bussa Amy S, Oleson Jacob, Claussen Alexander D, Gantz Bruce J, Hansen Marlan R, Scheperle Rachel A

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To evaluate the spatiotemporal dynamics of cochlear implant electrode impedance in standard and dexamethasone-eluting electrode arrays (Dex-EA).

STUDY

DESIGN

Prospective, double-arm study.

SETTING

Single tertiary referral center.

PATIENTS: Thirty-one (17 control and 14 experimental) subjects meeting Food and Drug Administration criteria for cochlear implantation.

INTERVENTIONS: Standard-of-care cochlear implant with intraoperative postinsertion impedance measurement.

MAIN OUTCOME MEASURES: Daily impedance measurements for a targeted 90 days postinitial activation.

RESULTS

This study demonstrates the spatiotemporal dynamics of electrode impedance observed over the first 90 days postimplantation with standard EAs, and the effects of Dex-EAs on these dynamics. For the control group, impedances for the 12 basal electrodes (1 to 12) increase over time while they decrease for the 10 apical electrodes (13 to 22). In comparison, not only do Dex-EAs result in smaller impedance values across all 22 electrodes over the first 90 days after initial activation, but the impedances also tend to decrease over time at all electrode locations. There is a statistically significant difference in impedance between the 2 cohorts at every electrode (P<0.0001), but the difference is largest for the basal electrodes. Dex-EAs also result in lower within-subject variability across repeated impedance measures over this short time window.

CONCLUSIONS

Dex-EAs result in smaller, more stable impedance values across repeated measures over the first 90 days after initial activation as compared with commercially available standard CI electrode arrays. The dramatically lower impedances, reduced fluctuations, and resistance to rising values particularly at basal locations reflect the clinical advantages of Dex-EA toward diminishing the foreign body response and may also enhance stimulation from the CI.

Для цитирования: Khan Uzair A, Bussa Amy S, Oleson Jacob, Claussen Alexander D, Gantz Bruce J, Hansen Marlan R, Scheperle Rachel A. Spatiotemporal Dynamics of Cochlear Implant Electrode Impedance in Standard and Dexamethasone-Eluting Electrode Arrays.. Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology. 2025. DOI: 10.1097/mao.0000000000004747.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.