DOI 10.1016/j.jelectrocard.2026.154424

Медицинская библиотека · Journal of electrocardiology · 2026

Physical reconfiguration of limb electrodes for Precordial Bipolar Lead acquisition: Morphological validation against digital subtraction.

Obregón-Rosas Santiago, Mc Loughlin Mario Jorge, Carlos-Solís Paola Fabiola, Vera-Ávila Nayra, Rojas-Bergés Isabella, Navarrete-Punaro José Pablo

СтатьяВалидационное исследованиеПеревод в очереди

Аннотация

BACKGROUND

The V2 - V1 Precordial Bipolar Lead (PBL) selectively evaluates the right-to-left retrosternal axis and has shown diagnostic value beyond the standard 12‑lead electrocardiogram. However, its use has been limited by the need for raw electrocardiographic data and post-processing software. This study evaluated whether a simple physical reconfiguration of limb electrodes could reproduce the digitally derived V2 - V1 morphology with sufficient accuracy for clinical application.

METHODS

Thirty-seven subjects underwent two sequential 10-s 12‑lead recordings using a Cardiovit FT-1 electrocardiograph sampled at 1000 Hz. In the standard recording, the digital PBL was calculated as V2 - V1. In the second recording, the right-arm and left-arm electrodes were repositioned to the V1 and V2 sites so that Lead I directly recorded the retrosternal dipole. Signals were filtered, synchronized, and analyzed using median beats. Morphological agreement was assessed with Pearson correlation on Z-normalized signals, while absolute agreement was evaluated using Lin's concordance correlation coefficient (CCC), intraclass correlation coefficient (ICC (Lewis, 1931; Nehb, 1938 [1,2])), root mean square error (RMSE), and Bland-Altman analysis.

RESULTS

Mean Pearson correlation between digital and physical PBL was 0.955 (SD 0.043), with segment-specific correlations of 0.953 (SD 0.054) for QRS and 0.967 (SD 0.052) for ST-T. Lin's CCC and ICC(2,1) were both 0.871 (SD 0.110), and RMSE was 0.091 (SD 0.049) mV. Bland-Altman analysis showed minimal bias (-0.008 mV).

CONCLUSIONS

Physical acquisition of the V2 - V1 PBL achieved high agreement with the digitally derived signal, supporting a simplified analog method for broader clinical implementation.

Для цитирования: Obregón-Rosas Santiago, Mc Loughlin Mario Jorge, Carlos-Solís Paola Fabiola, Vera-Ávila Nayra, Rojas-Bergés Isabella, Navarrete-Punaro José Pablo. Physical reconfiguration of limb electrodes for Precordial Bipolar Lead acquisition: Morphological validation against digital subtraction.. Journal of electrocardiology. 2026. DOI: 10.1016/j.jelectrocard.2026.154424.