van der Meijden Olivier A, Wijdicks Coen A, Gaskill Trevor R, Jansson Kyle S, Millett Peter J

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

PURPOSE: The purpose of this study was to compare single-row (SR), extended double-row (DR), and augmented, extended double-row (aDR) rotator cuff repairs in a two-tendon, posterosuperior rotator cuff tear (RCT) model with intact rotator cuff tendons.

METHODS

RCTs were created and randomly assigned to SR, DR, or aDR repair (5 each) in 20 cadaveric shoulder specimens. A collagen scaffold was used for augmentation. In the remaining 5 specimens, the rotator cuffs were left intact. All specimens were cyclically loaded from 25 to 75 N for 50 cycles. Every 50 cycles, peak load was increased by 25 N until failure occurred. Cyclic stiffness and number of cycles were analyzed.

RESULTS

The SR (72.9 ± 4.64 N/mm)- and aDR (72.6 ± 11.8 N/mm)-repaired specimens differed significantly in stiffness from the intact specimens (93.1 ± 14.8 N/mm) at ≥100 N (P < .05). The intact specimens and DR- and aDR-repaired specimens endured more cycles to failure (1,556 ± 677, 1,302 ± 248, and 1,211 ± 95, respectively) than the SR-repair specimens (388 ± 72 cycles, 260 ± 4 N) (P < .05 for all groups).

CONCLUSIONS

Linked DR constructs were significantly stronger than SR repairs in this two-tendon RCT model and approached the strength of the intact rotator cuff. Augmentation with a collagen patch (aDR) did not influence biomechanical repair qualities in this model, but did result in less variability in failure load and more consistency in the mode of failure.

CLINICAL RELEVANCE: The biomechanical properties of extended linked DR constructs are superior to those of SR constructs for repair of two-tendon RCTs, and are not compromised by graft augmentation.

Для цитирования: van der Meijden Olivier A, Wijdicks Coen A, Gaskill Trevor R, Jansson Kyle S, Millett Peter J. Biomechanical analysis of two-tendon posterosuperior rotator cuff tear repairs: extended linked repairs and augmented repairs.. Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association. 2013. DOI: 10.1016/j.arthro.2012.07.012.