Premature failure of small-sized Shelhigh No-React porcine pulmonic valve conduit model NR-4000.

Ishizaka Toru, Ohye Richard G, Goldberg Caren S, Ramsburg Stephen R, Suzuki Takaaki, Devaney Eric J, Bove Edward L

Оценочное исследованиеСтатьяПеревод в очереди

Аннотация

OBJECTIVE

Given the limited availability of small-sized cryopreserved pulmonary homografts, we implanted a series of Shelhigh No-React porcine pulmonic valve conduits (SPVC). The aim of this study was to evaluate the short-term performance following implantation.

METHODS

From February 2000 to September 2000, the SPVC was implanted 25 times in 24 patients in the right ventricular outflow tract (RVOT) to correct congenital anomalies. The anatomical malformations were TOF/PA in eight patients, TGA/VSD/PS in four, truncus arteriosus in four, IAA/VSD/AS or AA in four, l-TGA/VSD in two and other in two. Age at operation was 2.8+/-3.9 years (mean+/-SD), including 12 patients under 1 year. The median conduit size was 14mm (range, 10-18).

RESULTS

At a mean follow-up of 23+/-5 months, two late deaths (8%) have occurred. Although they were not primarily conduit related, both showed severe conduit stenosis. Twenty-one conduits (84%) showed mild to severe conduit stenosis, regurgitation or both. Two patients underwent balloon dilatation for distal conduit stenosis. Twelve conduits (48%) in 11 patients were removed at a median of 12 months (range, 2-18 months) due to RVOT obstruction in 11 and free conduit insufficiency with pseudoaneurysm in one. The typical findings of the explanted conduits were prominent intimal peel formation at the distal anastomosis without calcification. The actuarial freedom from reintervention at 18 months was 48+/-10%.

CONCLUSIONS

Our experience of the SPVC with the diameter of 14mm or less has revealed a high incidence of distal conduit stenosis due to intimal peel formation resulting in early conduit failure. These findings have led us to abandon its use when other options are available.

Для цитирования: Ishizaka Toru, Ohye Richard G, Goldberg Caren S, Ramsburg Stephen R, Suzuki Takaaki, Devaney Eric J, Bove Edward L. Premature failure of small-sized Shelhigh No-React porcine pulmonic valve conduit model NR-4000.. European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery. 2003. DOI: 10.1016/s1010-7940(03)00080-0.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.