OBJECTIVE

This study was undertaken to investigate the degeneration and calcification of valved bovine jugular vein segments for right ventricular outflow tract reconstruction in juvenile sheep.

METHODS

Seven valved bovine jugular vein conduits (Contegra model 220; VenPro Corporation, Irvine, Calif) and 3 control conduits (MH100; Medtronic, Inc, Minneapolis, Minn) were implanted in the pulmonary artery in young sheep. After 20 weeks the conduits were explanted and qualitatively analyzed by epicardial echocardiography, gross examination, x-ray analysis, light microscopy, and transmission electron microscopy. Calcification was determined quantitatively by flame atomic absorption spectrometry.

RESULTS

Two Contegra conduits could not be analyzed because of endocarditis. All other Contegra conduits functioned well, with preserved structure and minimal calcification. The control MH100 conduits exhibited extensive fibrous sheathing, with calcification of the aortic wall portion and the commissural part of the Hancock valve.

CONCLUSIONS

The Contegra conduit's performance was clearly superior to that of the control MH100 conduit when implanted in the pulmonary artery position in juvenile sheep for 5 months.

Для цитирования: Herijgers Paul, Ozaki Shigeyuki, Verbeken Erik, Van Lommel Alfons, Meuris Bart, Lesaffre Emmanuel, Daenen Willem, Flameng Willem. Valved jugular vein segments for right ventricular outflow tract reconstruction in young sheep.. The Journal of thoracic and cardiovascular surgery. 2002. DOI: 10.1067/mtc.2002.121043.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.