DOI 10.1016/j.jvscit.2023.101370
Emergent decommissioning of HeartMate 3 for life-threatening hemorrhage.
Аннотация
Left ventricular assist devices are known for the treatment of heart failure. We present a patient who underwent emergent decommissioning of his left ventricular assist device after experiencing life-threatening hemorrhage. In this case, the patient's outflow graft was avulsed from his left subclavian artery. We used a covered stent to first exclude the outflow graft and an Amplatzer II plug (Abbott Cardiovascular) to then occlude the inflow cannula. Accessing the inflow cannula of the left ventricular assist device was challenging. In hindsight, we could have accessed and occluded the outflow graft via the left subclavian artery before deploying the covered stent.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.