Медицинская библиотека · Cardiology in the young · 2024

Oxygen saturation measurement in cyanotic heart disease with the Apple watch.

Weis Angelika, Leroy Martin, Jux Christian, Rupp Stefan, Backhoff David

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Accurate measurement of transcutaneous oxygen saturation is important for the assessment of cyanosis in CHD. Aim of this study was the evaluation of a supplementary transcutaneous oxygen saturation measurement with an Apple watch® in children with cyanotic heart disease.

MATERIAL AND

METHODS

During a six-minute walk test, measurement of transcutaneous oxygen saturation was performed simultaneously with an Oximeter (Nellcor, Medtronic, USA) and an Apple watch® Series 7 (Apple inc, USA) in 36 children with cyanotic heart disease.

RESULTS

Median age was 9.2 (IQR 5.7-13.8) years. Transcutaneous oxygen saturation measurement with the Apple watch® was possible in 35/36 and 34/36 subjects before and after six-minute walk test. Children, in whom Apple watch® measurement was not possible, had a transcutaneous oxygen saturation < 85% on oximeter. Before six-minute walk test, median transcutaneous oxygen saturation was 93 (IQR 91-97) % measured by oximeter and 95 (IQR 93-96) % by the Apple watch®. After a median walking distance of 437 (IQR 360-487) m, transcutaneous oxygen saturation dropped to 92 (IQR 88-95, p < 0.001) % by oximeter and to 94 (IQR 90-96, p = 0.013) % measured with the Apple watch®.

CONCLUSION

In children with mild cyanosis measurement of transcutaneous oxygen saturation with an Apple watch® showed only valid results if transcutaneous oxygen saturation was > 85%, with higher values being measured with the smart watch. In children with moderate or severe cyanosis transcutaneous oxygen saturation, measurement with the Apple watch® was not reliable and cannot be recommended to monitor oxygen saturation at home.

Для цитирования: Weis Angelika, Leroy Martin, Jux Christian, Rupp Stefan, Backhoff David. Oxygen saturation measurement in cyanotic heart disease with the Apple watch.. Cardiology in the young. 2024. DOI: 10.1017/s1047951124025216.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.