Are orthopedic surgeons able to detect 3 mm and 5 mm joint step off on fluoroscopy?
Аннотация
OBJECTIVE
The aim of this study was to examine whether joint step-off created experimentally at 3 mm and 5 mm in the tibial lateral plateau can be accurately evaluated by orthopedic surgeons on fluoroscopic images.
PATIENTS AND
METHODS
A lateral tibia plateau fracture was created experimentally on above-the-knee amputated material. Using a ruler, step-off at 3 mm and then at 5 mm was made on the joint surface, then joint and lateral fluoroscopy images were obtained. These images were evaluated by 316 orthopedic surgeons. The surgeons were asked whether the joint congruence in the plateau fracture required surgical correction. The same question was asked again after 3 months, and all the responses were recorded.
RESULTS
In the first measurements for 3 mm joint step-off, 77 (24.4%) orthopedic surgeons stated that surgical correction was necessary, and for 5 mm, 118 (37.3%) surgeons stated that surgical correction was necessary. In the 3rd month, the need for surgical correction was stated by 144 (45.6%) surgeons for 3 mm, and by 176 (55.7%) surgeons for 5 mm (p=0.001)
CONCLUSIONS
Fluoroscopy is not a reliable method to determine articular step-off. Methods such as arthroscopic or open joint surface evaluation should be applied in the operating room.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.