MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00270-012-0411-3

Медицинская библиотека · Cardiovascular and interventional radiology · 2012

Unexpected complication with the new C3 excluder: cause and treatment.

Katsargyris Athanasios, Oikonomou Kyriakos, Bracale Umberto M, Verhoeven Eric L G

Клинический случайСтатьяПеревод в очереди

Аннотация

The new C3 Gore Excluder delivery system enables both up/downward and rotational repositioning of the device before complete deployment. This contributes to more precise proximal landing and permits facilitation of the contralateral gate cannulation. During separate deployment, the position of the ipsilateral limb can also be readjusted. We have used the modified C3 delivery system in more than 50 patients, and in most cases, we were able to utilize the repositioning options of the device to achieve optimal fixation and sealing. However, we present a case where our attempt to readjust the position of the ipsilateral limb led to upward migration of the main body with coverage of the left renal artery origin. The latter was secured with a bare stent implantation.

Для цитирования: Katsargyris Athanasios, Oikonomou Kyriakos, Bracale Umberto M, Verhoeven Eric L G. Unexpected complication with the new C3 excluder: cause and treatment.. Cardiovascular and interventional radiology. 2012. DOI: 10.1007/s00270-012-0411-3.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.