A neonatal in-vitro study on the effect of the inflation pressure on end-tidal carbon dioxide levels.

Dassios Theodore, Kaltsogianni Ourania, Saka Jonathan, Greenough Anne

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Describing the association of the peak inflation pressure (PIP) with end-tidal carbon dioxide (ETCO2) is a prerequisite for the development of closed loop ventilation in neonatal intensive care. We aimed to develop an in-vitro system to study this relationship.

METHODS

A ventilator was connected to a test lung, supplied with a stable CO2 concentration from a cylinder. The PIP was altered and the change in ETCO2 per unit of PIP was calculated in three models mimicking respiratory distress syndrome (RDS), bronchopulmonary dysplasia (BPD) and viral bronchiolitis.

RESULTS

The median (IQR) change in ETCO2 per unit of PIP was 0.23(0.13-0.38) kPa/cmH2O, using 138 paired measurements of PIP and ETCO2. The median (IQR) change in ETCO2 per unit of PIP, was higher when starting at an ETCO2 > 6 kPa [0.43(0.33-0.58) kPa/cmH2O] compared to starting at an ETCO2 < 6 kPa [0.14(0.08-0.20) kPa/cmH2O, p < 0.001]. The median (IQR) change in ETCO2 per unit of PIP, was larger in the model of RDS [0.33(0.13-0.51) kPa/cmH2O] compared to the BPD [0.23(0.13-0.33) kPa/cmH2O, p = 0.043] and the bronchiolitis models [0.15(0.10-0.31) kPa/cmH2O, p = 0.017].

CONCLUSIONS

The change in ETCO2 in response to increasing PIP was larger for higher ETCO2 values and in a model simulating neonatal RDS, compared to BPD and bronchiolitis.

Для цитирования: Dassios Theodore, Kaltsogianni Ourania, Saka Jonathan, Greenough Anne. A neonatal in-vitro study on the effect of the inflation pressure on end-tidal carbon dioxide levels.. Medical engineering & physics. 2023. DOI: 10.1016/j.medengphy.2023.104052.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.