MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcmg.2018.12.022

Медицинская библиотека · JACC. Cardiovascular imaging · 2019

Automated Pixel-Wise Quantitative Myocardial Perfusion Mapping by CMR to Detect Obstructive Coronary Artery Disease and Coronary Microvascular Dysfunction: Validation Against Invasive Coronary Physiology.

Kotecha Tushar, Martinez-Naharro Ana, Boldrini Michele, Knight Daniel, Hawkins Philip, Kalra Sundeep, Patel Deven, Coghlan Gerry, Moon James, Plein Sven, Lockie Tim, Rakhit Roby, Patel Niket, Xue Hui, Kellman Peter, Fontana Marianna

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАВалидационное исследованиеПеревод в очереди

Аннотация

OBJECTIVES

This study sought to assess the performance of cardiovascular magnetic resonance (CMR) myocardial perfusion mapping against invasive coronary physiology reference standards for detecting coronary artery disease (CAD, defined by fractional flow reserve [FFR] ≤0.80), microvascular dysfunction (MVD) (defined by index of microcirculatory resistance [IMR] ≥25) and the ability to differentiate between the two.

BACKGROUND

Differentiation of epicardial (CAD) and MVD in patients with stable angina remains challenging. Automated in-line CMR perfusion mapping enables quantification of myocardial blood flow (MBF) to be performed rapidly within a clinical workflow.

METHODS

Fifty patients with stable angina and 15 healthy volunteers underwent adenosine stress CMR at 1.5T with quantification of MBF and myocardial perfusion reserve (MPR). FFR and IMR were measured in 101 coronary arteries during subsequent angiography.

RESULTS

Twenty-seven patients had obstructive CAD and 23 had nonobstructed arteries (7 normal IMR, 16 abnormal IMR). FFR positive (epicardial stenosis) areas had significantly lower stress MBF (1.47 ± 0.48 ml/g/min) and MPR (1.75 ± 0.60) than FFR-negative IMR-positive (MVD) areas (stress MBF: 2.10 ± 0.35 ml/g/min; MPR: 2.41 ± 0.79) and normal areas (stress MBF: 2.47 ± 0.50 ml/g/min; MPR: 2.94 ± 0.81). Stress MBF ≤1.94 ml/g/min accurately detected obstructive CAD on a regional basis (area under the curve: 0.90; p < 0.001). In patients without regional perfusion defects, global stress MBF <1.82 ml/g/min accurately discriminated between obstructive 3-vessel disease and MVD (area under the curve: 0.94; p < 0.001).

CONCLUSIONS

This novel automated pixel-wise perfusion mapping technique can be used to detect physiologically significant CAD defined by FFR, MVD defined by IMR, and to differentiate MVD from multivessel coronary disease. A CMR-based diagnostic algorithm using perfusion mapping for detection of epicardial disease and MVD warrants further clinical validation.

Для цитирования: Kotecha Tushar, Martinez-Naharro Ana, Boldrini Michele, Knight Daniel, Hawkins Philip, Kalra Sundeep, Patel Deven, Coghlan Gerry, Moon James, Plein Sven, Lockie Tim, Rakhit Roby, Patel Niket, Xue Hui, Kellman Peter, Fontana Marianna. Automated Pixel-Wise Quantitative Myocardial Perfusion Mapping by CMR to Detect Obstructive Coronary Artery Disease and Coronary Microvascular Dysfunction: Validation Against Invasive Coronary Physiology.. JACC. Cardiovascular imaging. 2019. DOI: 10.1016/j.jcmg.2018.12.022.