Efficacy and safety of the ZSI375 artificial urinary sphincter for male stress urinary incontinence: lessons learned.

Kretschmer Alexander, Hüsch Tanja, Thomsen Frauke, Kronlachner Dominik, Pottek Tobias, Obaje Alice, Anding Ralf, Rose Achim, Olianas Roberto, Friedl Alexander, Hübner Wilhelm, Homberg Roland, Pfitzenmaier Jesco, Grein Ulrich, Queissert Fabian, Naumann Carsten M, Schweiger Josef, Wotzka Carola, Nyarangi-Dix Joanne N, Hofmann Torben, Buchner Alexander, Haferkamp Axel, Bauer Ricarda M, Debates On Male Incontinence (DOMINO)-Project

СтатьяМногоцентровое исследованиеПеревод в очереди

Аннотация

PURPOSE: To analyze efficacy and safety for the ZSI375 artificial urinary sphincter in a multicenter case series.

METHODS

Thirteen male patients with stress urinary incontinence underwent implantation of a ZSI375 artificial urinary sphincter device between 2010 and 2012 in three international continence reference centers. Perioperative characteristics and postoperative complications were analyzed using the Clavien-Dindo scale. Re-hospitalization and explantation rates, and functional outcome were assessed. Inner-group and between-group differences were analyzed using Wilcoxon, Mann-Whitney U, and Fisher's exact test whenever indicated. Kaplan-Meier analysis was performed to assess device survival. A p value below 0.05 was considered statistically significant.

RESULTS

There were no intraoperative complications. Median follow-up was 13.5 months. In this period, four device defects (30.8 %) could be observed, being the main cause for device explantation, followed by device infection (15.4 %), non-resolvable pain (7.7 %), and urethral erosion (7.7 %). There were no Clavien IV or Clavien V complications. Overall explantation rate was 61.5 %. Mean time-to-explantation was 279 ± 308 days. There was no significant influence of previous irradiation and previous invasive incontinence therapy (p = 0.587 and p = 0.685, respectively). Mean daily pad usage decreased from 5.8 ± 1.5 to 2.4 ± 2.1 (p = 0.066). One patient (7.7 %) did not use any pads. Social continence (0-1 pads) was achieved in 15.4 % of the patients.

CONCLUSION

This is the most current study that is investigating the outcome after ZSI375 implantation in a multicenter case series. Based on our results, explantation rates after ZSI375 implantation are high and efficacy rates seem lower than previously described. Addressing this high failure rate, the system has undergone a two-step modification in the meantime.

Для цитирования: Kretschmer Alexander, Hüsch Tanja, Thomsen Frauke, Kronlachner Dominik, Pottek Tobias, Obaje Alice, Anding Ralf, Rose Achim, Olianas Roberto, Friedl Alexander, Hübner Wilhelm, Homberg Roland, Pfitzenmaier Jesco, Grein Ulrich, Queissert Fabian, Naumann Carsten M, Schweiger Josef, Wotzka Carola, Nyarangi-Dix Joanne N, Hofmann Torben, Buchner Alexander, Haferkamp Axel, Bauer Ricarda M, Debates On Male Incontinence (DOMINO)-Project. Efficacy and safety of the ZSI375 artificial urinary sphincter for male stress urinary incontinence: lessons learned.. World journal of urology. 2016. DOI: 10.1007/s00345-016-1787-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.