Isoflurane anesthesia and body temperature changes do not significantly affect Viscoelastic Coagulation Monitor Vet viscoelastography of New Zealand white rabbits (Oryctolagus cuniculus).
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Doden Greta, Guthrie Daneeya, Rudra Pratyaydipta, McCready Julianne E, Brandão João
Аннотация
OBJECTIVE: To evaluate the impact of anesthesia and body temperature changes on a point-of-care viscoelastic coagulation monitoring device (VCM-Vet, Entegrion Corp) parameters in New Zealand white rabbits (Oryctolagus cuniculus).
METHODS: Purpose-bred New Zealand white rabbits were anesthetized in a research facility as part of a randomized, complete crossover study. Each rabbit was anesthetized using isoflurane twice (7 days apart), with or without a heated anesthesia circuit (Heated ZDS Qube; Darvall), over approximately 2 weeks. Lateral saphenous venipuncture was performed within 5 minutes of anesthetic induction (baseline) and at 50 minutes after anesthetic induction (T50) during each event. Fresh whole blood was immediately tested using a single VCM-Vet analyzer, providing 2 tracings per event. Statistical analysis was performed using a linear mixed model with significance at P < .05.
RESULTS: 10 approximately 4-month-old intact male apparently healthy rabbits were included. All rabbits were normothermic (median, 39.3 °C; range, 38.5 to 39.8 °C) at baseline, and all rectal temperatures were lower at T50 compared to baseline (median, 1.6 °C; range, 0.7 to 2.1 °C). There was no statistically significant difference in any clot parameter between baseline and T50 nor between different rectal temperatures; however, there were statistical differences in 4 clot parameters between nonheated and heated anesthetic events.
CONCLUSIONS: Anesthesia and mildly decreased body temperature did not significantly impact the viscoelastography assessment.
CLINICAL RELEVANCE: Although thromboelastography reference intervals have been established in rabbits, the use of the VCM-Vet analyzer is understudied in this species. Future research is needed to determine VCM-Vet reference intervals for rabbits.
METHODS: Purpose-bred New Zealand white rabbits were anesthetized in a research facility as part of a randomized, complete crossover study. Each rabbit was anesthetized using isoflurane twice (7 days apart), with or without a heated anesthesia circuit (Heated ZDS Qube; Darvall), over approximately 2 weeks. Lateral saphenous venipuncture was performed within 5 minutes of anesthetic induction (baseline) and at 50 minutes after anesthetic induction (T50) during each event. Fresh whole blood was immediately tested using a single VCM-Vet analyzer, providing 2 tracings per event. Statistical analysis was performed using a linear mixed model with significance at P < .05.
RESULTS: 10 approximately 4-month-old intact male apparently healthy rabbits were included. All rabbits were normothermic (median, 39.3 °C; range, 38.5 to 39.8 °C) at baseline, and all rectal temperatures were lower at T50 compared to baseline (median, 1.6 °C; range, 0.7 to 2.1 °C). There was no statistically significant difference in any clot parameter between baseline and T50 nor between different rectal temperatures; however, there were statistical differences in 4 clot parameters between nonheated and heated anesthetic events.
CONCLUSIONS: Anesthesia and mildly decreased body temperature did not significantly impact the viscoelastography assessment.
CLINICAL RELEVANCE: Although thromboelastography reference intervals have been established in rabbits, the use of the VCM-Vet analyzer is understudied in this species. Future research is needed to determine VCM-Vet reference intervals for rabbits.