Analytical chemistry · 2025
Integrated Analytical System for Comprehensive Two-Dimensional Enantio-Gas Chromatography and Low-Pressure Gas Chromatography Mass Spectrometry Utilizing a Switching Valve: Design and Optimization.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Rinaldi Giorgia, Ferracane Antonio, Zoccali Mariosimone, Mondello Luigi
Аннотация
An analytical system was developed to perform both comprehensive two-dimensional enantio-gas chromatography mass spectrometry with the second-dimension column working under low-pressure conditions (eGC×LP-GC-QMS) and low-pressure gas chromatography mass spectrometry (LP-GC-QMS) analysis, without any changes to the instrumental setup. This is achieved using a switching valve, which also allows for back-flushing of the first-dimension column, thereby generating a flexible and powerful analytical platform. The eGC×LP-GC-QMS configuration can provide both targeted and untargeted information in a single run. In this research, it is employed for the analysis of 56 pesticides (without considering isomers), with the added capability of the backflush mode, effectively removing high-boiling compounds extracted from the matrix and improving sample throughput and system robustness, without major concerns regarding sample preparation. The first-dimension is a chiral column, which enables the separation of seven enantiomers: acephate, benoxacor, chlorflurecol, dibrom, fipronil, methamidophos, and propetamphos. The second-dimension is a wide-bore column (5 m × 0.53 mm I.D. × 0.53 μm df) operating under vacuum conditions. Thanks to the use of a switching valve, it is possible to directly perform LP-GC-QMS analysis, enabling the investigation of a broader range of compounds in terms of boiling point. The operability range of volatile analytes was evaluated for both configurations by injecting different alkane mixtures: C7-C30 for the eGC×LP-GC-QMS configuration and C7-C40 for the LP-GC-QMS configuration.