Results of the multicentre evaluation of an electrochemiluminescence immunoassay for HCG on Elecsys 2010.
Аннотация
This study evaluated the performance of the HCG STAT Elecsys assay in 8 European laboratories using the Elecsys 2010 system. Analytical sensitivity was < 0.5 mlU/mL. The analysis of concentration series prepared by mixing serum pools with high and low HCG concentrations proved linearity up to 10.000 mIU/mL. A high-dose hook effect was not seen up to HCG concentrations of 430.000 mIU/mL. The medians of the within-run CVs (n = 21, 3 series) were 3.0% (2.1-5.8% CV; 10.4-14.4 mIU/mL), 2.4% (1.7-6.1% CV; 35.6-88.6 mIU/mL) and 2.3% (1.7-6.1% CV; 282.3-643.8 mIU/mL). The medians of the between-day imprecisions (n = 10-21) were 7.0% CV (5.2-12.0% CV; 4.0-14.0 mIU/mL), 5.5% CV (3.1-7.2% CV; 35.4-92.7% mIU/mL) and 4.1% CV (2.8-5.1% CV; 270.8-658.0 mIU/mL). The median recovery of two external quality control samples with assigned values of 9.39 and 10.40 mIU/mL) were 101.2 and 104.3% (ranges: 94.8-116.1%, 98.6-117.8%, n = 10). The assay was compared with five non-isotopic automated routine immunoassay systems (x). Slopes ranged from 0.87 to 1.15 and intercepts from-0.53 to 12.50 mIU/mL. The coefficients of correlation were with one exception (0.898) > or = 0.960. The distribution of HCG in samples from non-pregnant women and healthy men was very similar to that observed with other automated routine methods. The HCG Elecsys assay is very specific for the intact holo-hormone. Nicked HCG dimer, nicked and non-nicked beta-subunits are weakly recognised or not detected. Hemoglobin, bilirubin and lipemia (tested up to: Hb, 3.7 g/L; bilirubin, 500 mumol/L; triglyceride, 37.6 mmol/L) did not interfere the assay. The HCG Elecsys assay is well suited for the early and fast diagnosis of normal pregnancy and the detection of tubal pregnancy.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.