DOI 10.1213/xaa.0000000000002011
A Persistent Endotracheal Tube Cuff Leak Due to an Undiagnosed Tracheal Diverticulum: A Case Report.
Аннотация
An endotracheal tube (ETT) cuff leakage has serious consequences. Anatomical abnormalities are rare but are possible causes. We report the case of a patient having a robot-assisted laparoscopic prostatectomy. Endotracheal intubation was performed successfully. However, a cuff leak occurred, and persisted despite standard management efforts. The ETT was carefully advanced until one-lung ventilation was established and then gradually withdrawn until bilateral lung ventilation was restored. Chest computed tomography revealed a tracheal diverticulum immediately below the vocal cords as the underlying cause. If standard troubleshooting fails to resolve a cuff leak, unusual causes should be considered.