DOI 10.2214/ajr.17.18261

Медицинская библиотека · AJR. American journal of roentgenology · 2018

Lens Dose Reduction by Patient Posture Modification During Neck CT.

Mosher Elizabeth G, Butman John A, Folio Les R, Biassou Nadia M, Lee Choonsik

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

Radiation exposure of the lens during neck CT may increase a patient's risk of developing cataracts. Radiologists at the National Institutes of Health worked with technicians to modify the neck CT scanning procedure to include a reduction in the scanning range, a reduction in the tube potential (kilovoltage), and a change in neck positioning using a head tilt. We objectively quantified the organ dose changes after this procedure modification using a computer simulation.

MATERIALS AND

METHODS

We retrospectively analyzed CT images of 40 patients (20 men and 20 women) scanned before and after the procedure change. Radiation dose to the lens delivered before and after the procedure change was calculated using an in-house CT dose calculator combined with computational human phantoms deformed to match head tilt angles. We also calculated the doses to other radiosensitive organs including the brain, pituitary gland, eye globes, and salivary glands before and after the procedure change.

RESULTS

Our dose calculations showed that modifying the neck position, shortening the scanning range, and reducing the tube potential reduced the dose to the lens by 89% (p < 0.0001). The median brain, pituitary gland, globes, and salivary gland doses also decreased by 59%, 52%, 66%, and 29%, respectively. We found that overranging significantly affects the lens dose.

CONCLUSION

Combining head tilt and scanning range reduction is an easy and effective method that significantly reduces radiation dose to the lens and other radiosensitive head and neck organs.

Для цитирования: Mosher Elizabeth G, Butman John A, Folio Les R, Biassou Nadia M, Lee Choonsik. Lens Dose Reduction by Patient Posture Modification During Neck CT.. AJR. American journal of roentgenology. 2018. DOI: 10.2214/ajr.17.18261.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.