Pressure-derived estimations of coronary flow reserve are inferior to flow-derived coronary flow reserve as diagnostic and risk stratification tools.
Аннотация
BACKGROUND
Pressure-derived coronary flow reserve (CFRpres) and pressure-bounded CFR (CFRpb) enable simple estimation of CFR from routine pressure measurements, but have been inadequately validated. We sought to compare CFRpres and CFRpb against flow-derived CFR (CFRflow) in terms of diagnostic accuracy, as well as regarding their comparative prognostic relevance.
METHODS
We evaluated 453 intermediate coronary lesions with intracoronary pressure and flow measurements. CFR was defined as hyperemic flow/baseline flow. The lower bound (CFRpres) and upper bound of CFRpb were defined as √[(ΔPhyperemia) / (ΔPrest)] and [(ΔPhyperemia) / (ΔPrest)], respectively. Long-term follow-up (median: 11.8-years) was performed in 153 lesions deferred from treatment to document the occurrence of major adverse cardiac events (MACE) defined as a composite of cardiac death, myocardial infarction and target vessel revascularization. CFR < 2.0 was considered abnormal.
RESULTS
CFRpb was normal or abnormal in 56.7% of stenoses, and indeterminate in 43.3% of stenoses. There was a poor diagnostic agreement between CFRpres and CFRpb with CFRflow (overall agreement: 45.5% and 71.6% of vessels, respectively). There was equivalent risk for long-term MACE for lesions with abnormal versus normal CFRpres (Breslow p = 0.562), whereas vessels with abnormal CFRflow were significantly associated with increased long-term MACE (Breslow p < 0.001). For vessels where CFRpb was abnormal or normal, there was equivalent risk for long-term MACE for vessels with abnormal versus normal CFRpb (Breslow p = 0.194), whereas vessels with abnormal CFRflow were associated with increased MACE rates over time (Breslow p < 0.001).
CONCLUSIONS
Pressure-derived estimations of CFR poorly agree with flow-derived measurements of CFR, which may explain the inferior association with long-term MACE as compared to flow-derived CFR.
Предметные рубрики MeSH