DOI 10.1186/s12882-023-03359-z

Медицинская библиотека · BMC nephrology · 2023

Non-contrast low-dose CT can be used for volumetry of ADPKD.

Yoo Jaeyeong, Kim Jin Up, Kim Jisu, Jeon Sohyun, Song Young-Jin, Choi Kwang-Ho, Kim Seok-Hyung, Yoon Jong-Woo, Kim Hyunsuk

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Kidney volume provides important information for the diagnosis and prognosis of autosomal dominant polycystic kidney disease (ADPKD), as well as for the evaluation of the effects of drugs such as tolvaptan. Non-contrast computed tomography (CT) is commonly used for volumetry, and this study examined the correspondence and correlation of kidney volume measured by standard-dose or low-dose CT.

METHODS

Axial standard-dose and low-dose CT images with 1-mm slices were obtained from 24 ADPKD patients. The kidney was segmented in the Synapse 3D software and the kidney volume was calculated using stereology. The kidney volume was compared between the two sets of images using R2, Bland-Altman plots, coefficient of variation, and intra-class correlation coefficients (ICCs).

RESULTS

The mean age of the 24 patients was 48.4 ± 10.9 years, and 45.8% were men (n = 11). The mean total kidney volume on standard-dose CT was 1501 ± 838.2 mL. The R2 of volume between standard-dose and low-dose CT was 0.995. In the Bland-Altman plot, except for one case with a large kidney volume, the two measurements were consistent, and the coefficient of variation and ICC were also good (0.02, 0.998). The CT radiation dose (dose-length product) was 229 ± 68 mGy·cm for standard-dose CT and 50 ± 19 mGy·cm for low-dose CT. A comparable volume was obtained with 20% of the radiation dose of standard-dose CT.

CONCLUSIONS

Standard-dose and low-dose CT showed comparable kidney volume in ADPKD. Therefore, low-dose CT can substitute for ADPKD volumetry while minimizing radiation exposure.

Для цитирования: Yoo Jaeyeong, Kim Jin Up, Kim Jisu, Jeon Sohyun, Song Young-Jin, Choi Kwang-Ho, Kim Seok-Hyung, Yoon Jong-Woo, Kim Hyunsuk. Non-contrast low-dose CT can be used for volumetry of ADPKD.. BMC nephrology. 2023. DOI: 10.1186/s12882-023-03359-z.