Disinfection of needleless connectors to reduce Staphylococcus aureus bacterial load.
Аннотация
HIGHLIGHTS: Compare effectiveness of chemical disinfectants in reducing S. aureus. Five disinfectants reduced the bacterial load, especially chlorhexidine solutions. Focus on Brazilian clinical practice of needleless connector disinfection.
PURPOSE: This study aimed to gain further knowledge about the comparative effectiveness of chemical disinfectants in reducing the bacterial load of NCs inoculated with S. aureus.
METHODS
Disinfection of needleless connectors was undertaken in vitro against S. aureus comparing 70% isopropyl alcohol (IPA), 70% ethanol, 0.5% and 2% chlorhexidine in 70% IPA applied with gauze, and 70% IPA single-use cap (Site-Scrub®).
RESULTS
All disinfectants reduced the bacterial load (P<0.001), especially the chlorhexidine solutions. Mechanical friction should follow guidelines.
CONCLUSION
This study found that all tested disinfectants effectively reduced the bacterial load and more clinical studies must be developed with a focus on the Brazilian clinical practice of needleless connector disinfection.