DOI 10.12968/bjon.2022.31.19.s26

Медицинская библиотека · British journal of nursing (Mark Allen Publishing) · 2022

Disinfection of needleless connectors to reduce Staphylococcus aureus bacterial load.

Rocha Patrícia Kuerten, Rickard Claire M, Gales Ana Cristina, Sincero Thaís Cristine Marques, Ray-Barruel Gillian, Ullman Amanda J, Dalcin Camila Biazus, Pedreira Mavilde L Gonçalves

СтатьяПеревод в очереди

Аннотация

HIGHLIGHTS: Compare effectiveness of chemical disinfectants in reducing S. aureus. Five disinfectants reduced the bacterial load, especially chlorhexidine solutions. Focus on Brazilian clinical practice of needleless connector disinfection.

PURPOSE: This study aimed to gain further knowledge about the comparative effectiveness of chemical disinfectants in reducing the bacterial load of NCs inoculated with S. aureus.

METHODS

Disinfection of needleless connectors was undertaken in vitro against S. aureus comparing 70% isopropyl alcohol (IPA), 70% ethanol, 0.5% and 2% chlorhexidine in 70% IPA applied with gauze, and 70% IPA single-use cap (Site-Scrub®).

RESULTS

All disinfectants reduced the bacterial load (P<0.001), especially the chlorhexidine solutions. Mechanical friction should follow guidelines.

CONCLUSION

This study found that all tested disinfectants effectively reduced the bacterial load and more clinical studies must be developed with a focus on the Brazilian clinical practice of needleless connector disinfection.

Для цитирования: Rocha Patrícia Kuerten, Rickard Claire M, Gales Ana Cristina, Sincero Thaís Cristine Marques, Ray-Barruel Gillian, Ullman Amanda J, Dalcin Camila Biazus, Pedreira Mavilde L Gonçalves. Disinfection of needleless connectors to reduce Staphylococcus aureus bacterial load.. British journal of nursing (Mark Allen Publishing). 2022. DOI: 10.12968/bjon.2022.31.19.s26.