Impact of Volume Guarantee on High-Frequency Oscillatory Ventilation in Preterm Infants: A Randomized Crossover Clinical Trial.

Iscan Burcin, Duman Nuray, Tuzun Funda, Kumral Abdullah, Ozkan Hasan

Сравнительное исследованиеСтатьяРандомизированное исследованиеПеревод в очереди

Аннотация

BACKGROUND

High-frequency oscillatory ventilation (HFOV) with volume guarantee (VG) is a new ventilation mode that allows the clinician to set a mean tidal volume to be delivered.

OBJECTIVE

This study aimed to investigate whether HFOV with a VG option may result in constant tidal volume delivery and less fluctuant CO2 levels compared to HFOV alone in premature infants with respiratory distress syndrome (RDS).

METHODS

Inborn infants at less than 32 weeks of gestation with RDS requiring invasive mechanical ventilation were eligible. Patients were randomized to receive HFOV + VG or HFOV alone as the initial ventilator mode and then crossed over to the other mode. HFOV was performed with 'optimal lung volume strategy' during both of the periods.

RESULTS

Twenty infants were evaluated. The mean high-frequency tidal volume (VThf) and CO2 diffusion coefficient (DCO2) were significantly higher in the HFOV + VG mode than HFOV alone. HFOV + VG maintains VThf within the target range more consistently than HFOV. The incidences of hypocarbia and hypercarbia were lower in HFOV with VG than HFOV alone.

CONCLUSION

This is the first prospective, randomized, short-term crossover clinical study that compared HFOV with and without VG in infants with acute RDS. Because of the lower VThf fluctuation and lower incidences of out-of-target PCO2 levels, HFOV combined with VG seems to be feasible for preterm infants. However, the results should be interpreted with caution due to the small sample size and short-term crossover design of the study.

Для цитирования: Iscan Burcin, Duman Nuray, Tuzun Funda, Kumral Abdullah, Ozkan Hasan. Impact of Volume Guarantee on High-Frequency Oscillatory Ventilation in Preterm Infants: A Randomized Crossover Clinical Trial.. Neonatology. 2015. DOI: 10.1159/000437204.