Patient-specific dose estimation considering subject thickness with an RGB-D camera: A phantom study.

Yasue K, Fuse H, Nakazawa Y, Takahashi T, Nosaka H, Koori N, Takahashi M, Miyakawa S, Hanada K, Urikura A, Fujisaki T, Abe S

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

Patient-specific radiation dose management in radiography is essential for optimizing diagnostic quality while minimizing patient exposure. However, current methods often rely on standardized protocols or empirical judgment, which may be inaccurate for patients with non-standard body sizes. Accurate, non-invasive measurement of subject thickness is critical for estimating entrance surface dose (ESD) tailored to each patient. This study aimed to evaluate the feasibility and accuracy of using an RGB-D camera for real-time subject thickness measurement and patient-specific ESD estimation in abdominal radiography.

METHODS

An abdominal phantom with five thicknesses (20-40 cm) was constructed using an anthropomorphic model and water-equivalent slabs. An RGB-D camera (Intel RealSense D435i) was mounted on a radiography unit to acquire depth measurements at the detector center and four off-center positions (5 cm upward, downward, leftward, and rightward). Measured thicknesses were compared to caliper measurements. ESD was calculated using both RGB-D and caliper-derived values. Levene's test assessed positional effects on measurement variability.

RESULTS

At the detector center, RGB-D-derived thicknesses deviated by no more than ±0.3 cm from actual values. Off-center measurements showed slightly larger errors, up to ±1.2 cm. Significant variance differences were found only at 20 cm thickness in the lateral positions (p < 0.05). ESD estimates from RGB-D-based thickness measurements differed by no more than 0.4 mGy, with percentage differences within ±0.7 %.

CONCLUSION

RGB-D camera-based thickness measurement is a feasible and accurate method for non-invasive, patient-specific ESD estimation in abdominal radiography.

IMPLICATIONS FOR PRACTICE: This approach enables pre-imaging dose estimation, supports personalized dose control, and may improve safety and consistency in radiographic practice.

Для цитирования: Yasue K, Fuse H, Nakazawa Y, Takahashi T, Nosaka H, Koori N, Takahashi M, Miyakawa S, Hanada K, Urikura A, Fujisaki T, Abe S. Patient-specific dose estimation considering subject thickness with an RGB-D camera: A phantom study.. Radiography (London, England : 1995). 2025. DOI: 10.1016/j.radi.2025.103181.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.