DOI 10.1007/s10840-020-00843-6
Technical considerations for CRT-D implantation in different varieties of persistent left superior vena cava.
Аннотация
PURPOSE: The persistent left superior vena cava (PLSVC) is usually asymptomatic and creates a challenge when detected incidentally during cardiac resynchronization therapy defibrillator (CRT-D) implantation. The purpose of our cases is to show different anatomical variables of PLSVC and different strategies used for CRT-D implantation.
METHODS
Four cases of PLSVC were presented. Pre-procedural bilateral venography was done to define anatomical variant of PLSVC. The side of approach and vein of approach were chosen according to the anatomical variant. Major challenges, electrical parameters, procedural times, long-term follow up, and complications were addressed.
RESULTS
Two cases were de novo CRT-D implantation. One case was an extraction/re-implantation of the coil lead, and one case was an upgrading. In one case, CRT-D implantation was followed by AVN ablation. All cases had successful devices implantation. Two cases had isolated PLSVC: one of them had right approach and the other had left approach. One case had double SVC with no connecting brachiocephalic veins and underwent a left-sided approach. One case had double SVC with a small connecting brachiocephalic vein and had a left approach for implantation with using the small brachiocephalic vein for the RV lead. Electrical parameters were acceptable for all leads implanted. Long-term follow-up was done for 6 months to 5 years. One complication occurred (acute atrial lead dislodgement).
CONCLUSIONS
In our case series, the presence of PLSVC did not preclude successful placement of pacemaker/defibrillator leads using standard tools. Bilateral venography helped to decide the side and vein of lead insertion.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.