Comparison of echocardiographic and fluoroscopic sizing of the left atrial appendage prior to percutaneous closure.

Kiani Soroosh, Patel Akshar Vipul, El-Chami Mikhael F, Patel Anshul M, Merchant Faisal M, Westerman Stacy B, De Lurgio David B, Hoskins Michael H

СтатьяПеревод в очереди

Аннотация

PURPOSE: Sizing of the left atrial appendage (LAA) ostium prior to occlusion (LAAO) is routinely performed with transesophageal echocardiography (TEE). We sought to compare the utility of sizing via fluoroscopy to TEE for percutaneous LAAO.

METHODS

We retrospectively evaluated all patients undergoing percutaneous LAAO at our institution from April 2015 through January 2018 (n = 195). We evaluated baseline characteristics, maximum measured ostium size (for both TEE and fluoroscopy), and differences in measured size to device size for both techniques.

RESULTS

Of the total cohort, 185 (95%) had both intraoperative TEE and fluoroscopic images available for analysis and were included in the final analysis. The mean age was 74 years and 64% were male. Hypertension was present in 89%, diabetes in 30%, and stroke in 32% of patients. The mean ejection fraction was 51%. The most common LAA morphology was "wind sock" (52%). Measured ostial diameter on fluoroscopy was larger compared with TEE (2.04 ± 3.43 mm larger on fluoroscopy, p < 0.001). Ostium diameter on TEE was more closely correlated to the size of the device implanted compared with fluoroscopy (0.76 vs. 0.61, p = 0.001).

CONCLUSIONS

Fluoroscopy results in larger estimated LAA ostium diameter compared with TEE. Despite this, TEE was more strongly correlated to operator choice in device sizing, which may reflect practice patterns. Because compression of the ostium on the device is necessary for long-term procedural success, under-sizing may lead to a higher rate of leaks. Prospective evaluation of the utility of routine fluoroscopic sizing compared with TEE is warranted.

Для цитирования: Kiani Soroosh, Patel Akshar Vipul, El-Chami Mikhael F, Patel Anshul M, Merchant Faisal M, Westerman Stacy B, De Lurgio David B, Hoskins Michael H. Comparison of echocardiographic and fluoroscopic sizing of the left atrial appendage prior to percutaneous closure.. Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing. 2019. DOI: 10.1007/s10840-019-00643-7.