MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/js9.0000000000004501

Медицинская библиотека · International journal of surgery (London, England) · 2025

Robotic augmentation of surgical skills in subretinal injection: a novel semiautomated system.

Lian Zhangkai, Xu Andi, Li Yanlin, Liu Jiali, Song Rihui, Liang Hongli, Wang Ruixin, Yang Yahan, Wei Xiaoyue, Zhang Zheming, Shang Yuanjun, Iao Wai Cheng, Jin Ling, Zhao Lanqin, Liu Bingqian, Ma Wei, Li Tao, Chen Wenben, Ke Qiong, Yan Pisong, Wu Xiaohang, Lin Duoru, Nasseri Mohammad Ali, Huang Kai, Lin Haotian

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Subretinal injection (SRI) is a microsurgical procedure requiring micron-level precision and stable hand control, where technical difficulty and reliance on surgeon expertise often compromise consistency and safety. We developed a semiautomated high-precision ophthalmic surgical robot (HiPOSuR) to enhance precision and reduce variability.

MATERIALS AND

METHODS

A total of 48 pigmented rabbits were randomized into manual (n = 24) and robot-assisted (n = 24) groups. Each procedure was performed by two retinal specialists and four residents. The primary outcome was the first-attempt success rate of SRI. Secondary outcomes included overall success rate, intraoperative complications, surgical duration, injection site healing, retinal recovery, and tremor amplitude based on video analysis. Surgeon perceptions were evaluated via post-procedure questionnaires.

RESULTS

The robot-assisted group demonstrated a first-attempt success rate of 91.7% (22/24), significantly surpassing the manual group's success rate of 37.5% (9/24; P < 0.001, 95% CI: 30.6%-77.7%). Intraoperative complications decreased from 54.2% (manual) to 8.3% (robot-assisted; P < 0.001, 95% CI: - 69.9% to -21.7%). Notably, robot-assisted residents achieved an 83.3% success rate (vs. 33.3% manually; P = 0.036, 95% CI: 12.4%- 87.6%), with complications decreasing from 58.3% to 16.7% (P = 0.003, 95% CI: - 99.6% to -33.7%). In the robot-assisted group, average tremor amplitude was markedly reduced (34.56 ± 18.75 μm vs. 95.32 ± 46.92 μm, P < 0.001) and questionnaire scores were significantly higher across all 10 items (P < 0.05).

CONCLUSION

HiPOSuR improves subretinal injection by minimizing tremor, and reducing dependence on surgeon experience. It offers a promising platform for robotic assistance in complex intraocular procedures.

Для цитирования: Lian Zhangkai, Xu Andi, Li Yanlin, Liu Jiali, Song Rihui, Liang Hongli, Wang Ruixin, Yang Yahan, Wei Xiaoyue, Zhang Zheming, Shang Yuanjun, Iao Wai Cheng, Jin Ling, Zhao Lanqin, Liu Bingqian, Ma Wei, Li Tao, Chen Wenben, Ke Qiong, Yan Pisong, Wu Xiaohang, Lin Duoru, Nasseri Mohammad Ali, Huang Kai, Lin Haotian. Robotic augmentation of surgical skills in subretinal injection: a novel semiautomated system.. International journal of surgery (London, England). 2025. DOI: 10.1097/js9.0000000000004501.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.