MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1186/s12245-026-01312-z

Медицинская библиотека · International journal of emergency medicine · 2026

An exploratory feasibility study of a novel portable mainstream capnograph in a prehospital environment.

Pedrotti Niccolò, Larsson Anders, Fagerlund Malin Jonsson, Smekal David

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Monitoring end-tidal carbon dioxide (EtCO2) and respiratory rate (RR) is a non-invasive and reliable, continuous breathing assessment method in intubated and spontaneously breathing patients. The prehospital spreading of this technique in awake patients has been limited by mainstream capnographs´ dimensions and costs. Yet the potential prehospital applications of a wearable mainstream capnograph are many. This study aims to assess the ability to measure EtCO2 of the novel mainstream MARIE capnograph in normal ambulance transport conditions.

METHODS

A single-centre, observational, exploratory feasibility study conducted in Region Uppsala between December 2024 and January 2025. First 20 healthy adult volunteers were transported approximately 10 min by ambulance, having EtCO2 and RR continuously measured by the mainstream device. Three measurements were performed (in the beginning, in the middle and at the end of the transport). A fourth value was noted on arrival in the Emergency Department (ED) with a reference capnograph for comparison with the third value by the mainstream device. The same measurement procedure was then repeated on 30 clinically stable adult patients transported by ambulance to the ED.

RESULTS

A total of 20 volunteers and 29 patients were included in the final analysis. The comparison between the mainstream and reference capnograph showed that median EtCO2 values were 4.6 [4.1-5.2] vs. 4.9 [4.4-5.2] kPa respectively (p = 0.35). Bland Altman analysis showed a mean difference between the mainstream capnograph and reference device of 0.16 kPa (95% CI -0.07-0.41), with moderate correlation between measurements. The median values of breaths per minute were 17 [14-21] vs. 19 [15-22] respectively (p < 0.01).

CONCLUSION

This study suggests that the mainstream capnograph MARIE is able to measure credible EtCO2 values in standard ambulance transport conditions.

CLINICAL TRIAL REGISTRATION: ClinicalTrials.gov, TRN NCT06905613, Registration date 16th December 2024.

Для цитирования: Pedrotti Niccolò, Larsson Anders, Fagerlund Malin Jonsson, Smekal David. An exploratory feasibility study of a novel portable mainstream capnograph in a prehospital environment.. International journal of emergency medicine. 2026. DOI: 10.1186/s12245-026-01312-z.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.