MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jhin.2025.09.014

Медицинская библиотека · The Journal of hospital infection · 2025

Optimizing endoscope surveillance: how sampling methods and time points affect microbiological surveillance.

Xu T, Cao X, Shuang M, Tan Q, Xiong L

СтатьяМногоцентровое исследованиеПеревод в очереди

Аннотация

BACKGROUND

Endoscope-associated infections remain a persistent concern despite adherence to reprocessing guidelines. Microbial surveillance is essential for quality control, but its effectiveness may depend on sampling timing and technique.

METHODS

This prospective multi-centre study was conducted across three gastrointestinal endoscopy units within multiple campuses of a tertiary general hospital in Wuhan, China. Sampling was performed either immediately after high-level disinfection (HLD) or following overnight storage in a dry cabinet using one of two techniques: the conventional flushing sampling method (CFSM) or the flush-brush-flush sampling method (FBFSM).

RESULTS

The overall pass rate for disinfected digestive endoscopes was 83.4% (212/254). Compared with CFSM, FBFSM exhibited a significantly lower overall pass rate (χ2=5.895, P=0.015), indicating enhanced microbial recovery. Endoscopes sampled after overnight drying showed a significantly higher pass rate (87.7% vs 79.3%) and lower detection rates for high-concern organisms, particularly Pseudomonas spp. and Gram-positive cocci, compared with samples collected immediately after HLD.

CONCLUSIONS

The drying phase of endoscope reprocessing plays a critical role in reducing residual microbial contamination. Additionally, FBFSM demonstrated superior sensitivity in post-reprocessing surveillance, highlighting its value as a more effective sampling method compared with CFSM.

Для цитирования: Xu T, Cao X, Shuang M, Tan Q, Xiong L. Optimizing endoscope surveillance: how sampling methods and time points affect microbiological surveillance.. The Journal of hospital infection. 2025. DOI: 10.1016/j.jhin.2025.09.014.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.