BACKGROUND

Tapered, fluted titanium (TFT) femoral stems have become the gold standard in revision total hip arthroplasty (rTHA). However, there is a paucity of data on TFT stem subsidence rates following aseptic rTHA. Subsidence can lead to instability, mechanical failure, leg-length discrepancy, and may require revision surgery. This study evaluated the incidences and predictors of TFT subsidence in aseptic rTHA.

METHODS

A total of 102 TFT femoral stems of 4 designs were retrospectively reviewed. Stem subsidence was measured on digital radiographs taken immediately after surgery and at standard clinical follow-up. Patient characteristics, risk factors for subsidence, revision etiologies, and implant characteristics were recorded. Patient-reported outcome measures were also evaluated for a subset of cases.

RESULTS

Overall, 12% of stems subsided >1 cm, and subsidence was minimal (<3 mm) in ≥64% of cases. From immediate postoperative to 1-month radiographic follow-up, 79% of stems subsided a mean of 2.9 mm (range, 0.1 to 12 mm). Beyond 1 month, subsidence was minimal for ≥77% of cases. In multivariate analyses, women and less femoral implant canal fill were associated with greater subsidence (P ≤ .034). The TFT stem design was not associated with early subsidence (P = .816). There were no modular junction fractures. There were 2 fractures and 2 subsidence-related revisions for aseptic loosening that occurred postoperatively.

CONCLUSIONS

The amount of subsidence in TFT stems was low and was detectable in the early (less than 1 year) postoperative period. Maximizing TFT stem fill within the femoral canal appears to reduce the risk of subsidence without increasing femoral fracture rates and should be the goal with implantation of these devices.

LEVEL OF EVIDENCE: IV-Case Series, No Control Group.

Для цитирования: Baldwin Thomas J, Deckard Evan R, Buller Leonard T, Meneghini R Michael. Incidence and Predictors of Subsidence Using Modular, Tapered, Fluted Titanium Femoral Stems in Aseptic Revision Total Hip Arthroplasty.. The Journal of arthroplasty. 2023. DOI: 10.1016/j.arth.2023.10.057.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.