OBJECTIVE

To evaluate one novel and two previously reported Doppler flow velocimetric techniques to estimate atrioventricular (AV) time intervals, suggested to be useful for early identification of fetuses at risk for congenital heart block.

METHODS

In 22 newborn infants, Doppler tracings were obtained from the mitral valve/aortic outflow and the superior vena cava/ascending aorta, as an ECG was recorded simultaneously. AV time intervals were measured using the onsets of the mitral A-wave/aortic outflow (MV-Ao), superior vena cava a-wave/aortic flow (SVC-Ao), and mitral A-wave/mitral valve closure (MV) as indirect markers of electrical atrial/ventricular activation.

RESULTS

Close positive linear relationships to the electrocardiographic PR interval were demonstrated for the MV-Ao (r = 0.82, S(y/x) = 7.4 ms), SVC-Ao (r = 0.85, S(y/x) = 6.8 ms), and MV (r = 0.92, S(y/x) = 3.8 ms) approaches. Both techniques using the aortic flow to indicate ventricular activation overestimated the PR interval: the MV-Ao by + 32 +/- 7.7 ms (mean +/- SD) and the SVC-Ao approach by + 22 +/- 7.0 ms. The new MV approach using mitral closure for the same purpose did not overestimate the PR interval, but there was a trend towards underestimation of the PR intervals as time intervals increased.

CONCLUSIONS

When systematic differences between echocardiographic and electrocardiographic AV time intervals are compensated for, all three techniques are useful to get indirect estimates of the PR interval. As MV recordings only need insonation of a single valve, and are thus easier to obtain, this technique may be of value as a first screening method to identify fetuses in need for further surveillance. In cases with AV time prolongation the SVC-Ao method seems superior.

Для цитирования: Bergman G, Jacobsson L-A, Wahren-Herlenius M, Sonesson S-E. Doppler echocardiographic and electrocardiographic atrioventricular time intervals in newborn infants: evaluation of techniques for surveillance of fetuses at risk for congenital heart block.. Ultrasound in obstetrics & gynecology : the official journal of the International Society of Ultrasound in Obstetrics and Gynecology. 2006. DOI: 10.1002/uog.2712.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.