Intermittent Pacemaker Malfunction Caused by Continuous Compression of the Lead by the Clavicle (Subclavian Crush Syndrome).

Rizal Ardian, Ruspiono Evit, Putri Dinarsari Hayuning

СтатьяПеревод в очереди

Аннотация

UNLABELLED: Subclavian vein access is still one of the most favoured access options for cardiac implantable electronic device (CIED) implantation. For the physician, the technique is reasonably familiar and easy to carry out. However, this has several potential complications. In this case, we present a late complication of subclavian access. The patient presented with intermittent loss of pacemaker output, which caused him to experience several syncopal events. In the acute setting, we changed the lead polarity and achieved a good outcome. Further management of this situation consisted of removal and replacement of the damaged lead.

LEARNING POINTS: Subclavian vein access is still one of the most favoured access options for cardiac device implantation.Intermittent loss of output can be a sign of pacemaker malfunction due to subclavian crush syndrome.Damaged lead extraction may be difficult to perform.

Для цитирования: Rizal Ardian, Ruspiono Evit, Putri Dinarsari Hayuning. Intermittent Pacemaker Malfunction Caused by Continuous Compression of the Lead by the Clavicle (Subclavian Crush Syndrome).. European journal of case reports in internal medicine. 2020. DOI: 10.12890/2020_001684.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.