Computer-assisted navigation applied to fetal cardiac intervention.
Аннотация
BACKGROUND
Prenatal cardiac interventions (PCI) for human fetal aortic valve (AoV) stenosis can reduce left ventricular hypoplasia and restore ventricular growth and function. However, 'freehand' needle delivery from the maternal skin through the uterine wall, fetal chest and ventricular apex to cross the fetal AoV remains technically challenging and time intensive, and is the rate-limiting step in the procedure.
METHODS
We developed a computer-assisted navigation (CANav) system that tracks the position and orientation of a two-dimensional (2D) ultrasound image relative to the trajectory of an electromagnetic (EM) embedded needle and stylet. We tested the CANav system in vitro using a water bath phantom, then in vivo using adult rats and pregnant (fetal) sheep.
RESULTS
The CANav system accurately tracked the delivered needle position in both in vitro phantom and adult rat model experiments. We performed 22 PCI attempts with or without CANav in a fetal sheep model. Maternal laparotomy was required to adjust the fetal position in 50% of the procedures. The time required to deliver the needle from the skin into the left ventricle (LV) using CANav was 2.9 +/- 1.7 (range 2-7) min (n = 14) vs. 5.5 +/- 4.3 (range 1-12) min (n = 8) without CANav (p < 0.05). The time needed to cross the aortic valve once the needle was within the LV was similar with and without CANav (p = 0.19).
CONCLUSIONS
CANav reduces the PCI time required to accurately deliver a needle to the fetal heart. Adaptations of this technical approach may be relevant to other congenital cardiac conditions and ultrasound-guided medical procedures.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.